Hamilton Medical AG recalls IntelliCuff
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Due to reported complaints, the cuff device may alarm with a "Cuff System Leakage" error. As a result, the motor may pump continuously and fail to maintain the cuff pressure set by the user.
- Products
-
- IntelliCuff
- Company
- Hamilton Medical AG
- Where it was sold
-
Worldwide - US Nationwide distribution in the states of AZ, CA, CO, FL, GA, HI, IL, KY, MA, MI, ND, NE, NE, NY, OH, PA, PR, SD, TX, VA, WA, WI, WV, WY and the country of Canada.
States: Arizona, California, Colorado, Florida, Georgia, Hawaii, Illinois, Kentucky, Massachusetts, Michigan, North Dakota, Nebraska, New York, Ohio, Pennsylvania, Puerto Rico, South Dakota, Texas, Virginia, Washington, Wisconsin, West Virginia, Wyoming
- Barcodes
07630002800839- Units affected
- 962 units
- About this recall
- IntelliCuff, Portable automatic cuff pressure controller REF: 951001