FDAMedical devicesRecalled Status: Ongoing

Hamilton Medical AG recalls Breathing circuit set

Class Iofficial

FDA Class I

FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.

Hazard
Ventilator coaxial breathing circuit may have a crack in blue tubing, which is used to deliver fresh gas to the patient. The crack will compromise inspiratory and expiratory flows, leading to partial or complete rebreathing of exhaled gases, which could lead to acute hypercapnia, respiratory acidosis, and if unrecognized organ dysfunction.
Products
  • Breathing circuit set
Where it was sold

US Nationwide distribution in the states of NH, NY, CT, PA, IL, ME, CO, NC, NJ, UT, TX, GA, SC, VA, IN, SD, MO, OR, TN, MA, AZ, NM.

States: Arizona, Colorado, Connecticut, Georgia, Illinois, Indiana, Massachusetts, Maine, Missouri, North Carolina, New Hampshire, New Jersey, New Mexico, New York, Oregon, Pennsylvania, South Carolina, South Dakota, Tennessee, Texas, Utah, Virginia

Barcodes
07630002802963
Units affected
2,560
About this recall
Breathing circuit set, coaxial, 1.80 m, with flow sensor and expiratory valve set, REF: 260128, used with HAMILTON-C1, HAMILTON-T1 and HAMILTON-MR1 ventilators.