Hamilton Medical AG recalls Breathing circuit set
FDA Class I
FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.
- Hazard
- Ventilator coaxial breathing circuit may have a crack in blue tubing, which is used to deliver fresh gas to the patient. The crack will compromise inspiratory and expiratory flows, leading to partial or complete rebreathing of exhaled gases, which could lead to acute hypercapnia, respiratory acidosis, and if unrecognized organ dysfunction.
- Products
-
- Breathing circuit set
- Company
- Hamilton Medical AG
- Where it was sold
-
US Nationwide distribution in the states of NH, NY, CT, PA, IL, ME, CO, NC, NJ, UT, TX, GA, SC, VA, IN, SD, MO, OR, TN, MA, AZ, NM.
States: Arizona, Colorado, Connecticut, Georgia, Illinois, Indiana, Massachusetts, Maine, Missouri, North Carolina, New Hampshire, New Jersey, New Mexico, New York, Oregon, Pennsylvania, South Carolina, South Dakota, Tennessee, Texas, Utah, Virginia
- Barcodes
07630002802963- Units affected
- 2,560
- About this recall
- Breathing circuit set, coaxial, 1.80 m, with flow sensor and expiratory valve set, REF: 260128, used with HAMILTON-C1, HAMILTON-T1 and HAMILTON-MR1 ventilators.