Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

GE Healthcare Finland Oy recalls CARESCAPE Canvas Smart Display

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
GE HealthCare has become aware of the potential for loss of monitoring on CARESCAPE Canvas 1000, CARESCAPE Canvas Smart Display, and certain versions of CARESCAPE B850 and CARESCAPE B650 monitors if, following the replacement of the CPU timekeeper battery, a mains power loss occurs during active monitoring. In the unlikely event that this occurs, it can result in the displays or monitors not turning on, after the mains power is restored. This can result in a delay in recognizing changes in patient condition until a replacement monitor is connected.
Products
  • CARESCAPE Canvas Smart Display
Where it was sold

Worldwide distribution.

Nationwide

Barcodes
00195278522313
Units affected
547 units
About this recall
CARESCAPE Canvas Smart Display, Model Numbers: a) 5514418, b) 5514418 / MAEP2, c) 5514418-01059642, d) 5514418-01080402, e) 5514418-01081317, f) 5514418-01086007, g) 5514418-01086838, h) 5514418-01088182, i) 5514418-01108442, j) 5514418-01118673, k) 5514418-01120764, l) 5514418-01166521, m) 5514418-01180241, n) 5514418-01181177, o) 5514418-01183037, p) 5514418-01184556, q) 5514418-01184557, r) 5514418-01185805, s) 5514418-01186277, t) 5514418-01187473, u) 5514418-01188453, v) 5514418-01190677, w) 5514418-01192703, x) 5514418-01192705, y) 5514418-01203242, z) 5514418-01206279, aa) 5514418-01208248, bb) 5514418-01212670, cc) 5514418-01215947, dd) 5514418-01218564, ee) 5514418-01222489, ff) 5514418-01222812, gg) 5514418-01223831, hh) ii) 5514418-01228038, ii) 5514418-01228181, jj) 5514418-01230534, kk) 5514418-01235322, ll) 5514418-01236388, mm) 5514418-01244357, nn) 5514418-01250690, oo) 5514418-01250761, pp) 5514418-01256439; monitor, physiological, patient