Siemens Medical Solutions USA, Inc recalls ARTIS Icono biplane
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- During patient examinations, the system may sporadically display a 0 x-ray dose administered to the patient. This behavior has only been observed during fluoroscopy and/or single-frame acquisitions. Despite the reported x-ray dose showing 0, an actual x-ray dose is applied to the patient.
- Products
-
- ARTIS Icono biplane
- Where it was sold
-
US and Worldwide
Nationwide
- Barcodes
405686906331740568692959234056869149325- About this recall
- ARTIS Icono biplane, floor and ceiling.: ARTIS icono biplane Model 11327600. ARTIS icono ceiling Model 11328100. ARTIS icono floor Model 11327700. interventional fluoroscopic x-ray system