GE Medical Systems Information Technologies Inc recalls GE HealthCare CARESCAPE TELEMETRY SERVER V5 ATO MODEL
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- GE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result from a persistent NO COMM or OFF NETWORK condition at the CARESCAPE Central Station or CIC Pro Clinical Information Center (Central Station).
- Products
-
- GE HealthCare CARESCAPE TELEMETRY SERVER V5 ATO MODEL
- Where it was sold
-
Worldwide distribution.
Nationwide
- Barcodes
00840682144773- Units affected
- 476 units
- About this recall
- GE HealthCare CARESCAPE TELEMETRY SERVER V5 ATO MODEL, REF #2063702-101, Product Code SS9; used in conjunction with ApexPro CARESCAPE Telemetry Server (CTS) v5.0; Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)