Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

GE Medical Systems Information Technologies Inc recalls GE HealthCare TRADE IN CARESCAPE TELEMETRY SERVER ARK-2250L - NORTH AMERICA

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
GE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result from a persistent NO COMM or OFF NETWORK condition at the CARESCAPE Central Station or CIC Pro Clinical Information Center (Central Station).
Products
  • GE HealthCare TRADE IN CARESCAPE TELEMETRY SERVER ARK-2250L - NORTH AMERICA
Where it was sold

Worldwide distribution.

Nationwide

Units affected
2366 units
About this recall
GE HealthCare TRADE IN CARESCAPE TELEMETRY SERVER ARK-2250L - NORTH AMERICA, REF #5697603-902, Product Code KSA; used in conjunction wtih ApexPro CARESCAPE Telemetry Server (CTS) v6.0; Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)