Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Datex-Ohmeda Inc recalls Giraffe Stand-alone Infant Resuscitation System; Powered emergency ventilator (resuscitator) Product identifie

Class Iofficial

FDA Class I

FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.

Hazard
Blender knob shaft can loosen, which can affect the delivered oxygen concentration. The system can deliver an oxygen concentration that does not match the intended concentration set by the user, potentially resulting in hypoxia or hyperoxia in the patient.
Products
  • Giraffe Stand-alone Infant Resuscitation System; Powered emergency ventilator (resuscitator) Product identifie
Where it was sold

Pending

Units affected
550 (367 US; 183 OUS)
About this recall
Giraffe Stand-alone Infant Resuscitation System; Powered emergency ventilator (resuscitator) Product identifiers: 2082663-001, 2082663-002, M1241424 Product impacted only if service blender M1091607-R was installed during the last Preventative Maintenance