Datex-Ohmeda Inc recalls Giraffe Warmer with Integrated Resuscitation System (iRes)
FDA Class I
FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.
- Hazard
- Blender knob shaft can loosen, which can affect the delivered oxygen concentration. The system can deliver an oxygen concentration that does not match the intended concentration set by the user, potentially resulting in hypoxia or hyperoxia in the patient.
- Products
-
- Giraffe Warmer with Integrated Resuscitation System (iRes)
- Company
- Datex-Ohmeda Inc
- Where it was sold
-
Pending
- Units affected
- 550 (367 US; 183 OUS)
- About this recall
- Giraffe Warmer with Integrated Resuscitation System (iRes), Panda Warmer with Integrated Resuscitation System (iRes) Product Identifiers: M1118179, M1112198 Impacted only if service blender M1091607-R was installed during the last Preventative Maintenance.