Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Datex-Ohmeda Inc recalls Panda Freestanding Warmer

Class Iofficial

FDA Class I

FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.

Hazard
Blender knob shaft can loosen, which can affect the delivered oxygen concentration. The system can deliver an oxygen concentration that does not match the intended concentration set by the user, potentially resulting in hypoxia or hyperoxia in the patient.
Products
  • Panda Freestanding Warmer
Where it was sold

Pending

Units affected
550 (367 US; 183 OUS)
About this recall
Panda Freestanding Warmer, Infant Radiant Warmer Product Identifiers: 2063568-001 Impacted only if service blender M1091607-R was installed during the last Preventative Maintenance.