GE Medical Systems, LLC recalls Giraffe OmniBed
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- GE HealthCare has become aware of a potential issue involving the canopy soffit beneath the control panel end of the device. Under certain conditions, for example, impact during transport, one or more of the fasteners that secure the soffit can disengage. If this occurs, a fastener could fall into the patient compartment. In rare circumstances, if multiple fasteners disengage, the canopy soffit can become loose or detach. These conditions could potentially result in patient injury.
- Products
-
- Giraffe OmniBed
- Company
- GE Medical Systems, LLC
- Where it was sold
-
Worldwide - US Nationwide distribution.
Nationwide
- Units affected
- 19587 units
- About this recall
- Giraffe OmniBed, a combination of an infant incubator and an infant warmer, All Models