Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

GE Medical Systems, LLC recalls Giraffe OmniBed Carestations and Giraffe OmniBed Carestation CS1

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
GE HealthCare has become aware of a potential issue involving the canopy soffit beneath the control panel end of the device. Under certain conditions, for example, impact during transport, one or more of the fasteners that secure the soffit can disengage. If this occurs, a fastener could fall into the patient compartment. In rare circumstances, if multiple fasteners disengage, the canopy soffit can become loose or detach. These conditions could potentially result in patient injury.
Products
  • Giraffe OmniBed Carestations and Giraffe OmniBed Carestation CS1
Where it was sold

Worldwide - US Nationwide distribution.

Nationwide

Units affected
23430 units
About this recall
Giraffe OmniBed Carestations and Giraffe OmniBed Carestation CS1, a combination of an infant incubator and an infant warmer, Model Numbers 2082844-001-XXX