Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Medtronic Perfusion Systems recalls Octopus Evolution Tissue Stabilizer

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
During the manufacturing process, Medtronic personnel identified an assembly issue where the tubing was incorrectly positioned. The short tube was attached to the bottom of the canister, and the long tube was attached to the top, contrary to the intended design.
Products
  • Octopus Evolution Tissue Stabilizer
Where it was sold

Worldwide - US Nationwide distribution.

Nationwide

Barcodes
00643169668911
Units affected
134 units
About this recall
Octopus Evolution Tissue Stabilizer, Model TS2000