Medtronic Perfusion Systems recalls Octopus 4 Tissue Stabilizer
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- During the manufacturing process, Medtronic personnel identified an assembly issue where the tubing was incorrectly positioned. The short tube was attached to the bottom of the canister, and the long tube was attached to the top, contrary to the intended design.
- Products
-
- Octopus 4 Tissue Stabilizer
- Company
- Medtronic Perfusion Systems
- Where it was sold
-
Worldwide - US Nationwide distribution.
Nationwide
- Barcodes
00763000543679- Units affected
- 590 units
- About this recall
- Octopus 4 Tissue Stabilizer, Model 29400