Abiomed, Inc. recalls Abiomed Impella CP Set with SmartAssist (10th Generation) containing affected 14Fr Low Profile Introducer Kits
FDA Class I
FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.
- Hazard
- Potential for thrombus formation during prolonged use of the introducer.
- Products
-
- Abiomed Impella CP Set with SmartAssist (10th Generation) containing affected 14Fr Low Profile Introducer Kits
- Company
- Abiomed, Inc.
- Where it was sold
-
Worldwide - US Nationwide distribution in the states of FL, IL, MN, NJ, NY, VA and the country of Japan.
States: Florida, Illinois, Minnesota, New Jersey, New York, Virginia
- Barcodes
0081350201356600813502013948- Units affected
- 168 units
- About this recall
- Abiomed Impella CP Set with SmartAssist (10th Generation) containing affected 14Fr Low Profile Introducer Kits. Product Codes: 1000413, 1000834.