Abiomed, Inc. recalls ABIOMED Impella CP Set
FDA Class I
FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.
- Hazard
- A potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers are provided in every pump set.
- Products
-
- ABIOMED Impella CP Set
- Company
- Abiomed, Inc.
- Where it was sold
-
Worldwide distribution. US Nationwide, Australia (AU), Austria (AT), Belgium (BE), Brazil (BR), Brunei (BN), Canada (CA), China (CN), Croatia (HR), Czech Republic (CZ), Denmark (DK), Finland (FI), France (FR), Germany (DE), Greece (GR), Hong Kong (HK), Ireland (IE), Israel (IL), Italy (IT), Japan (JP), Kuwait (KW), Luxembourg (LU), Malaysia (MY), Mexico (MX), Netherlands (NL), Norway (NO), Panama (PA), Poland (PL), Saudi Arabia (SA), Serbia (RS), Slovenia (SI), South Korea (KR), Spain (ES), Sweden (SE), Switzerland (CH), Taiwan (TW), Thailand (TH), United Arab Emirates (AE), and United Kingdom (GB).
Nationwide
- Barcodes
00813502012279- Units affected
- 87,277 units
- About this recall
- ABIOMED Impella CP Set, Pump Set Code 0048-0003, containing BIOMED Impella Introducer, 14 Fr, 13 cm & 25 cm, Product Code 0052-3046. heart pump and accessories