GE Medical Systems Information Technologies Inc recalls GE Healthcare MAC VU360
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Software issue issue with the MAC VU360 v3.0 system. After selecting the order refresh icon, an open preview or review screen might fail to close automatically after 2 minutes of inactivity. As a result, data from a previous patient may remain on the screen and could be incorrectly associated with a subsequent patient.
- Products
-
- GE Healthcare MAC VU360
- Where it was sold
-
Worldwide distribution - US Nationwide and the countries of Belgium, Hong Kong, Netherlands, Sweden, UK.
Nationwide
- Barcodes
00198953107263- Units affected
- 163 units
- About this recall
- GE Healthcare MAC VU360, REF# part number 2030361-023; Electrocardiograph