Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Siemens Healthcare Diagnostics, Inc. recalls Syva Emit tox Serum Tricyclic Antidepressant Calibrator/Controls SMN 10445390 Catalog No. 7C219UL

Class IIIofficial

FDA Class III

FDA classified this recall as Class III: the product is unlikely to cause health problems but breaks the agency's rules (a labeling mistake, for example).

Hazard
out-of-specification sodium azide concentration in Syva Emit Tox Serum Tricyclic Antidepressant Calibrator/Controls Lot U2. While the assay results were not impacted, the overall health risk to lab technicians due to this issue is low, as exposure is limited by small volumes and a 0.2% post-reconstitution sodium azide concentration, with the worst outcome being mild skin or eye irritation and no serious harm anticipated.
Products
  • Syva Emit tox Serum Tricyclic Antidepressant Calibrator/Controls SMN 10445390 Catalog No. 7C219UL
Where it was sold

Worldwide - US Nationwide distribution in the states CA, ID, IL, IN, MA, MS, NC, NM, NY, OH, OK, OR, PA, WA, and Northern Mariana Islands part of USA and the countries of Canada, South Africa, Brazil, Chile, Colombia.

Nationwide

Barcodes
00842768014321
Units affected
494 US; 114 OUS
About this recall
Syva Emit tox Serum Tricyclic Antidepressant Calibrator/Controls SMN 10445390 Catalog No. 7C219UL