Abiomed, Inc. recalls Pump 371 14F LT CMR Set; Product Code: 0048-0045;
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction.
- Products
-
- Pump 371 14F LT CMR Set; Product Code: 0048-0045;
- Company
- Abiomed, Inc.
- Where it was sold
-
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Belgium, Brazil, Brunei Darussalam, Canada, Croatia, Czeck Republic, Denmark, Finland, France, Germany, Hong Kong, India, Ireland, Israel, Italy, Japan, Kuwait, Luxembourg, Malaysia, Mexico, Netherlands, Norway, Panama, Poland, Saudi Arabia, Serbia, Singapore, Slovenia, Spain, Sweden, Switzerland, Taiwan, U.A.E., United Kingdom.
Nationwide
- Barcodes
00813502011876- Units affected
- 43,636 units
- About this recall
- Pump 371 14F LT CMR Set; Product Code: 0048-0045;