Siemens Healthcare Diagnostics, Inc. recalls Dimension Vista¿ Total Bilirubin (TBIL) Flex Reagent Cartridge; Siemens Material Number (SMN): 10445146;

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
During manufacturing, the last three tests from the 160 test flex of lot 23206BA (wells 8 or 10) were processed on the Dimension Vista platform. There was an issue with the filling line which led to some wells to underfill in lot 23206BA. Calibrator, quality control (QC) and patient results could have been impacted.
Products
  • Dimension Vista¿ Total Bilirubin (TBIL) Flex Reagent Cartridge; Siemens Material Number (SMN): 10445146;
Where it was sold

Domestic: Nationwide Distribution. International: Austria, Belgium, Canada, Czech Republic, Denmark, France, Germany, Italy, Qatar, Slovenia, Spain, Switzerland.

Nationwide

Barcodes
00842768015717
Units affected
5,371 units (4,394 US, 977 OUS)
About this recall
Dimension Vista¿ Total Bilirubin (TBIL) Flex Reagent Cartridge; Siemens Material Number (SMN): 10445146;