Siemens Healthcare Diagnostics, Inc. recalls ADVIA Centaur CA 19-9 (250 Test)-In vitro diagnostic use in the quantitative measurement of the CA 19-9 tumor-

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Positive bias for some samples around the CA 19-9 upper limit of normal of 35 U/m as compared to the previous lot. The sample bias observed was greatest in the Asia Pacific sample population. The observed bias is not proportional across the assay measuring interval and is less pronounced for sample results above a concentration of 110 U/mL.
Products
  • ADVIA Centaur CA 19-9 (250 Test)-In vitro diagnostic use in the quantitative measurement of the CA 19-9 tumor-
Where it was sold

Worldwide - US Nationwide distribution

Nationwide

Barcodes
00630414574028
About this recall
ADVIA Centaur CA 19-9 (250 Test)-In vitro diagnostic use in the quantitative measurement of the CA 19-9 tumor-associated antigen in human serum using the Atellica CI and Atellica IM Analyzers SMN: 10491244