Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Abiomed, Inc. recalls Impella Optical Controller

Class Iofficial

FDA Class I

FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.

Hazard
The potential for purge cassette recognition issues due to the failure of the purge flag component of the purge pressure sensor assembly.
Products
  • Impella Optical Controller
Where it was sold

Domestic: Nationwide Distribution. International: United Arab Emirates, Austria, Australia, Belgium, Brunei Darussalam, Brazil, Canada, Switzerland, Chile, China, Colombia, Costa Rica, Cyprus, Czechia, Germany, Denmark, Spain, Finland, France, United Kingdom, Greece, Hong Kong, Croatia, Hungary, Ireland, Israel, India, Italy, Japan, Kuwait, Luxembourg, Mexico, Malaysia, Netherlands, Norway, New Zealand, Panama, Pakistan, Poland, Serbia, Saudi Arabia, Sweden, Singapore, Slovenia, Thailand, Taiwan;

Nationwide

Barcodes
00813502012972008135020129580081350201221700813502011814008135020124840081350201140100813502013160
Units affected
4100 units (2785 US, 1315 OUS) 11 10
About this recall
Impella Optical Controller, Automated Impella Controller, Impella Connect Package with the below product descriptions and corresponding Product Codes. 1. Optical, AIC, Impella Connect, Packaged, AU; Product Code: 0042-0040-AU. 2. Optical, AIC, Impella Connect, Packaged, CA; Product Code: 0042-0040-CA. 3. Optical, AIC, Impella Connect, Packaged, EU; Product Code: 0042-0040-EU. 4. Optical, AIC, Impella Connect, Packaged, JP; Product Code: 0042-0040-JP. 5. Optical, AIC, Impella Connect, Packaged, UK; Product Code: 0042-0040-UK. 6. Optical, AIC, Impella Connect, Packaged, US; Product Code: 0042-0040-US. 7. Optical, AIC, Impella Connect, Packaged, KR; Product Code: 0042-0040-KR;