Class IIofficial
Mckesson Medical-Surgical Inc. Corporate Office recalls Abbott Rapid Dx N America LLC Reproductive Endocrinology Control Set McKesson Consult" Pregnancy (hCG) TestingFacility Critical Temperature Excursions on 08-JAN 2024 in Walk-In refrigerator due to mechanical failures, causing temperatures to fall below the validated range of 35.6 - 46.4 Degrees Fahrenheit (2 - 8 Degrees Celsius) for up to 60 minutes which could increase the probability of an incorrect test result or a delayed test result.
-
Class IIofficial
Mckesson Medical-Surgical Inc. Corporate Office recalls Steris Corporation Endo Cleaning Indicator VERIFY" RESITEST"Facility Critical Temperature Excursions on 08-JAN 2024 in Walk-In refrigerator due to mechanical failures, causing temperatures to fall below the validated range of 35.6 - 46.4 Degrees Fahrenheit (2 - 8 Degrees Celsius) for up to 60 minutes which could increase the probability of an incorrect test result or a delayed test result.
-
Class IIofficial
Mckesson Medical-Surgical Inc. Corporate Office recalls Abbott Point of Care Control Control i-STAT¿ Level 1 10 X 1.7 mLFacility Critical Temperature Excursions on 08-JAN 2024 in Walk-In refrigerator due to mechanical failures, causing temperatures to fall below the validated range of 35.6 - 46.4 Degrees Fahrenheit (2 - 8 Degrees Celsius) for up to 60 minutes which could increase the probability of an incorrect test result or a delayed test result.
-
Class IIofficial
Mckesson Medical-Surgical Inc. Corporate Office recalls Abbott Point of Care Control Control i-STAT¿ Level 3 1.7 mLFacility Critical Temperature Excursions on 08-JAN 2024 in Walk-In refrigerator due to mechanical failures, causing temperatures to fall below the validated range of 35.6 - 46.4 Degrees Fahrenheit (2 - 8 Degrees Celsius) for up to 60 minutes which could increase the probability of an incorrect test result or a delayed test result.
-
Class IIofficial
Covidien recalls Medtronic Invosmanufacturing defects to sensor cable connectors may result in error codes or non-functioning channel
-
Class IIofficial
Covidien recalls Medtronic Invosmanufacturing defects to sensor cable connectors may result in error codes or non-functioning channel
-
Class IIofficial
Silex Medical, LLC recalls Signature Laparoscopic InstrumentsInsulation on the tube did not extend fully to the proximal end of the shaft on the instrument, leaving an exposed (not insulated) area of the shaft closer to the rotator wheel.
-
Class IIofficial
Silex Medical, LLC recalls Signature Laparoscopic InstrumentsInsulation on the tube did not extend fully to the proximal end of the shaft on the instrument, leaving an exposed (not insulated) area of the shaft closer to the rotator wheel.
-
Class IIofficial
Silex Medical, LLC recalls Signature Laparoscopic InstrumentsInsulation on the tube did not extend fully to the proximal end of the shaft on the instrument, leaving an exposed (not insulated) area of the shaft closer to the rotator wheel.
-
Class IIofficial
Silex Medical, LLC recalls Signature Laparoscopic InstrumentsInsulation on the tube did not extend fully to the proximal end of the shaft on the instrument, leaving an exposed (not insulated) area of the shaft closer to the rotator wheel.
-
Class IIofficial
Silex Medical, LLC recalls Signature Laparoscopic InstrumentsInsulation on the tube did not extend fully to the proximal end of the shaft on the instrument, leaving an exposed (not insulated) area of the shaft closer to the rotator wheel.
-
Class IIofficial
Silex Medical, LLC recalls Signature Laparoscopic InstrumentsInsulation on the tube did not extend fully to the proximal end of the shaft on the instrument, leaving an exposed (not insulated) area of the shaft closer to the rotator wheel.
-
Class IIofficial
Silex Medical, LLC recalls Signature Laparoscopic InstrumentsInsulation on the tube did not extend fully to the proximal end of the shaft on the instrument, leaving an exposed (not insulated) area of the shaft closer to the rotator wheel.
-
Class IIofficial
Silex Medical, LLC recalls Signature Laparoscopic InstrumentsInsulation on the tube did not extend fully to the proximal end of the shaft on the instrument, leaving an exposed (not insulated) area of the shaft closer to the rotator wheel.
-
Class IIofficial
Dr Franz Koehler Chemie Gmbh recalls Custodiol HTK SolutionPotential for particles to be present in solution.
-
Class IIofficial
Maquet Medical Systems USA recalls BEQ-RF-32-USA RotaFlow Centrifugal Pump with BIOLINE CoatingNonconformities at their seal could potentially compromise the sterile barrier for affected products. Breach of sterile barrier could expose patients to pathogenic agents, resulting in the following potential harms: inflammation, infection, or sepsis.
-
Class IIofficial
Maquet Medical Systems USA recalls BO-RF-32-USA RotaFlow Centrifugal Pump with SOFTLINE CoatingNonconformities at their seal could potentially compromise the sterile barrier for affected products. Breach of sterile barrier could expose patients to pathogenic agents, resulting in the following potential harms: inflammation, infection, or sepsis.
-
Class IIofficial
Becton Dickinson & Co. recalls Sensi Disc Amoxicillin with Clavulanic Acid 20/10 ¿gThere is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
-
Class IIofficial
Becton Dickinson & Co. recalls BD BBL Sensi Disc Ampicillin - 10 ¿gThere is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
-
Class IIofficial
Becton Dickinson & Co. recalls BD BBL Sensi Disc Ampicillin with Sulbactam 10/10 ¿gThere is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
-
Class IIofficial
Becton Dickinson & Co. recalls BD BBL Sensi Disc Augmentin - 3¿gThere is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
-
Class IIofficial
Becton Dickinson & Co. recalls BD BBL Sensi Disc Azithromycin -15 ¿gThere is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
-
Class IIofficial
Becton Dickinson & Co. recalls BD BBL Sensi Disc Aztreonam - 30 ug¿There is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
-
Class IIofficial
Becton Dickinson & Co. recalls BD BBL Sensi Disc Cefaclor - 30 ug¿There is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
-
Class IIofficial
Becton Dickinson & Co. recalls BD BBL Sensi Disc Cefepime - 30 ug¿There is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
-
Class IIofficial
Becton Dickinson & Co. recalls BD BBL Sensi Disc Cefixime - 5 ug¿There is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
-
Class IIofficial
Becton Dickinson & Co. recalls BD BBL Sensi Disc Cefotaxime - 30 ug¿There is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
-
Class IIofficial
Becton Dickinson & Co. recalls BD BBL Sensi Disc Cefpodoxime - 10 ugThere is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
-
Class IIofficial
Becton Dickinson & Co. recalls BD BBL Sensi Disc Ceftaroline - 30 ugThere is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
-
Class IIofficial
Becton Dickinson & Co. recalls BD BBL Sensi Disc Ceftazidime- 30 ugThere is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
-
Class IIofficial
Becton Dickinson & Co. recalls BD BBL Sensi Disc Ceftriaxone- 30 ugThere is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
-
Class IIofficial
Becton Dickinson & Co. recalls BD BBL Sensi Disc Cefuroxime- 30 ugThere is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
-
Class IIofficial
Becton Dickinson & Co. recalls BD BBL Sensi Disc Chloramphenicol- 30 ugThere is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
-
Class IIofficial
Becton Dickinson & Co. recalls BD BBL Sensi Disc Ciprofloxacin- 5 ugThere is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
-
Class IIofficial
Becton Dickinson & Co. recalls BD BBL Sensi Disc Clarithromycin- 15 ugThere is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
-
Class IIofficial
Becton Dickinson & Co. recalls BD BBL Sensi Disc Doripenem- 15 ugThere is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
-
Class IIofficial
Becton Dickinson & Co. recalls BD BBL Sensi Disc Ertapenem- 10 ugThere is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
-
Class IIofficial
Becton Dickinson & Co. recalls BD BBL Sensi Disc Imipenem- 10 ugThere is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
-
Class IIofficial
Becton Dickinson & Co. recalls BD BBL Sensi Disc Levofloxacin- 5 ugThere is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
-
Class IIofficial
Becton Dickinson & Co. recalls BD BBL Sensi Disc Meropenem- 10 ugThere is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
-
Class IIofficial
Becton Dickinson & Co. recalls BD BBL Sensi Disc Minocycline - 30 ugThere is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
-
Class IIofficial
Becton Dickinson & Co. recalls BD BBL Sensi Disc Moxifloxacin- 5 ugThere is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
-
Class IIofficial
Becton Dickinson & Co. recalls BD BBL Sensi Disc Ofloxacin- 5 ugThere is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
-
Class IIofficial
Becton Dickinson & Co. recalls BD BBL Sensi Disc Piperacillin/Tazobactam- 100/10 ugThere is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
-
Class IIofficial
Becton Dickinson & Co. recalls BD BBL Sensi Disc Rifampin- 5 ugThere is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
-
Class IIofficial
Becton Dickinson & Co. recalls BD BBL Sensi Disc Sulfamethoxazole with Trimethoprim 23.75/1.25 ugThere is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
-
Class IIofficial
Becton Dickinson & Co. recalls BD BBL Sensi Disc Tetracycline 30 ugThere is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.
-
Class IIofficial
Microbiologics Inc recalls Microbiologics Enteric Pathogens Control Panel (Inactivated Pellet)The negative control was contaminated with one of the positive control pathogens (Norovirus). This would result in users getting a positive call for Norovirus when running the negative control.
-
Class IIofficial
Impulse Dynamics (usa) Inc recalls OPTIMIZER model CCM X11 implantable pulse generator (IPG) devices - Smart Mini and LiteOPTIMIZER devices may cease to deliver CCM therapy if the device incorrectly detects a charging error. This may cause patients to experience heart failure symptoms similar to before implantation of the device.
-
Class IIofficial
Olympus Corporation of the Americas recalls Hemostasis ProbeThe BICOAG Hemostasis Probe with a coaxial plug was mispackaged into the package and labeled as a BICOAG Hemostasis Probe with a fixed pin connector.