Thoratec Corp. recalls Thoratec HeartMate 3Potential for a leak path at the seal interface between the left ventricle assist system inflow cannula and the titanium apical cuff.
-
Class Iofficial
Thoratec Corp. recalls Thoratec HeartMate 3Potential for a leak path at the seal interface between the left ventricle assist system inflow cannula and the titanium apical cuff.
-
Class IIofficial
Alere San Diego, Inc. recalls Abbott Cholestech LDX Battery KitThere is a potential that the Cholestech LDX battery (an external battery pack available as an optional accessory to the Cholestech LDX Analyzer) may swell which could cause minor burns, cuts or abrasions, or low current electrical shock to the user.
-
Class IIofficial
MEDLINE INDUSTRIES, LP - Northfield recalls Centurion OB INSTRUMENT TRAYTrumpet Needle Guide ring can detach when excessive pressure is applied.
-
Class IIofficial
MEDLINE INDUSTRIES, LP - Northfield recalls Medline medical procedure kitsTrumpet Needle Guide ring can detach when excessive pressure is applied.
-
Class IIofficial
Ortho-Clinical Diagnostics, Inc. recalls VITROS Immunodiagnostic Products Anti HBs Calibrators. The VITROS Anti-HBs Calibrator Pack contains 3 X labellPotential for positively biased results due to signal reduction over shelf life.
-
Class IIofficial
Pacific Lasertech, LLC recalls 25-LHP-828These laser products fail to comply with 21 CFR 1040.10(f)(4) Key control which requires, Each laser system classified as a Class IIIb or IV laser product shall incorporate a key-actuated master control. The key shall be removable and the laser shall not be operable when the key is removed. , because while the laser is switched on, the key can be removed from the key switch. The correction is replacement with a key switch that captures the key when in the on position.
-
Class IIofficial
Pacific Lasertech, LLC recalls 25-LHP-928These laser products fail to comply with 21 CFR 1040.10(f)(4) Key control which requires, Each laser system classified as a Class IIIb or IV laser product shall incorporate a key-actuated master control. The key shall be removable and the laser shall not be operable when the key is removed. , because while the laser is switched on, the key can be removed from the key switch. The correction is replacement with a key switch that captures the key when in the on position.
-
Class IIofficial
Pacific Lasertech, LLC recalls N-LHP-928These laser products fail to comply with 21 CFR 1040.10(f)(4) Key control which requires, Each laser system classified as a Class IIIb or IV laser product shall incorporate a key-actuated master control. The key shall be removable and the laser shall not be operable when the key is removed. , because while the laser is switched on, the key can be removed from the key switch. The correction is replacement with a key switch that captures the key when in the on position.
-
Class IIofficial
Optikem International, Inc. recalls TRP Natural Eyes Allergy Eyes ReliefLack of Assurance of Sterility
-
Class IIofficial
Optikem International, Inc. recalls TRP Blur ReliefLack of Assurance of Sterility
-
Class IIofficial
Optikem International, Inc. recalls TRP Eye Twitching ReliefLack of Assurance of Sterility
-
Class IIofficial
Optikem International, Inc. recalls TRP Natural Eyes Dryness ReliefLack of Assurance of Sterility
-
Class IIofficial
Optikem International, Inc. recalls TRP Eye strain ReliefLack of Assurance of Sterility
-
Class IIofficial
Optikem International, Inc. recalls TRP Natural Eyes Floaters ReliefLack of Assurance of Sterility
-
Class IIofficial
Optikem International, Inc. recalls TRP Natural Eyes Aging Eye ReliefLack of Assurance of Sterility
-
Class IIofficial
Optikem International, Inc. recalls TRP Natural Eyes Pink Eye ReliefLack of Assurance of Sterility
-
Class IIofficial
Optikem International, Inc. recalls TRP Natural Eyes Red Eye ReliefLack of Assurance of Sterility
-
Class IIofficial
Optikem International, Inc. recalls hyalogic For Dry EyesLack of Assurance of Sterility
-
Class IIofficial
SSM Health Care St. Louis DBA SSM St. Clare Health Center recalls Phenylephrine HClLack of Assurance of Sterility: Firm did not perform process validation.
-
Class IIofficial
SSM Health Care St. Louis DBA SSM St. Clare Health Center recalls FentaNYL citrateLack of Assurance of Sterility: Firm did not perform process validation.
-
Class IIofficial
Vitony Bakery, Inc. recalls Vitony BakeryUndeclared allergens FD&C Yellow #5, Yellow #6
-
Class IIofficial
Vitony Bakery, Inc. recalls Vitony BakeryUndeclared allergens FD&C Yellow #5, Yellow #6
-
Class IIofficial
Aji Mori Sauce Corp recalls AJI MORI***CHILLIMAMI SAUCE***UMAMI BOMB***NET WT. 250ml. Packaged in a LDPE bottle. Refrigerate after openingPotential contamination with Clostridium botulinum
-
Class IIofficial
Becton Dickinson & Co. recalls BD Phoenix NMIC-306 Panel is a self-inoculating (when used with a BD Phoenix Automated Microbiology System) poProducts were found to have confirmed drug cross contamination prior to release may lead to incorrect, inadequate, or absent antibiotic treatment or delay in results
-
Class IIofficial
Becton Dickinson & Co. recalls BD Phoenix Yeast ID Panel is a self-inoculating (when used with a BD Phoenix Automated Microbiology System) poEsculin, a substrate used in other panels and normally in A16 well, was found in A17 well. A17 in yeast panels is supposed to be an empty well and is not a part of the normal ID evaluation. However, A17 is checked for fluorescence and if anything is found, the system will abort the panel with the message Fluorescent interference has been detected. Esculin is expected to trigger this fluorescence and test cancellation.
-
Class IIofficial
Beckman Coulter, Inc. recalls Beckman Coulter DxI 9000 Access Immunoassay Analyzer is a discrete photometric chemistry analyzer for clinicalThe reason for the recall is the pneumatic tubing used in the analytical pick-and-place (PnP) assembly cable may fully or partially disconnect from the flex cable assembly fittings during normal operation. The DxI 9000 Access Immunoassay Analyzer may generate a System Event log message that indicates PnP errors. Any cancelled tests will have an associated SYS flag. The analyzer may enter the Red-system status because of this issue. The analyzer also may not eject RVs, where unbound conjugate could potentially splash up on the sidewall of the RV, not being properly washed away, resulting in an artificially higher signal (RLU). The analyzer may not detect this splashing and continue with normal operation, which can cause false low or high erroneous results. Consequently, the analyzer may cancel in-progress tests. These errors could potentially cause delay reporting patient test results and patient treatments. The degree of the potential health hazard is dependent on the analytes that are potentially delayed, especially those analytes that would be used in the acute setting when treatment decisions are immediate. In this particular recall the highest risk analyte is troponin. Health consequences could include permanent impairment or possibly death if medical treatment decisions are delayed due to a delay in obtaining patient results. The issue was identified by Beckman Coulter Inc. manufacturing staff at their location in Brea California on February 29, 2024, during in-process acceptance inspection activities.
-
Class IIofficial
Permobil Inc recalls Permobil M3 Corpus Power Wheelchair with Heavy Duty Shock REF:120031This recall has been reported retroactively. The power wheelchair firmware lacks a speed restriction to maintain stability when the seat lift exceeds 50 mm, which may cause the user to fall from the wheelchair or the wheelchair to tip over,
-
Class IIofficial
Baxter Healthcare Corporation recalls Baxter Peri-Guard Repair Patch (legacy product codes): a) PC0404NBIOA new contraindication was added to the IFU which states that Peri-Guard Repair Patch should not be used in neurosurgery since the product endotoxin level may be higher than the allowable limit for cerebrospinal fluid-contacting devices and Supple Peri-Guard Repair Patch should not be used in neurosurgery since the product endotoxin level may be higher than the allowable limit for cerebrospinal fluid-contacting devices.
-
Class IIofficial
Baxter Healthcare Corporation recalls Baxter Supple Peri-Guard Repair Patch (legacy product codes): a) PC0404SNBIOA new contraindication was added to the IFU which states that Peri-Guard Repair Patch should not be used in neurosurgery since the product endotoxin level may be higher than the allowable limit for cerebrospinal fluid-contacting devices and Supple Peri-Guard Repair Patch should not be used in neurosurgery since the product endotoxin level may be higher than the allowable limit for cerebrospinal fluid-contacting devices.
-
Class IIofficial
Baxter Healthcare Corporation recalls Baxter Peri-Guard Repair Patch (new product codes): a) PG0404A new contraindication was added to the IFU which states that Peri-Guard Repair Patch should not be used in neurosurgery since the product endotoxin level may be higher than the allowable limit for cerebrospinal fluid-contacting devices and Supple Peri-Guard Repair Patch should not be used in neurosurgery since the product endotoxin level may be higher than the allowable limit for cerebrospinal fluid-contacting devices.
-
Class IIofficial
Baxter Healthcare Corporation recalls Baxter Supple Peri-Guard Repair Patch (new product codes): a) SPG0404A new contraindication was added to the IFU which states that Peri-Guard Repair Patch should not be used in neurosurgery since the product endotoxin level may be higher than the allowable limit for cerebrospinal fluid-contacting devices and Supple Peri-Guard Repair Patch should not be used in neurosurgery since the product endotoxin level may be higher than the allowable limit for cerebrospinal fluid-contacting devices.
-
Class IIofficial
Philips North America recalls Patient Information Center iX (PIC iX)It was found that the Care Assist mobile application (version 4.1.1 and earlier) would not play custom ringtones when iOS device screens were locked with a passcode.
-
Class IIofficial
Philips North America recalls CareEvent Event Management SystemIt was found that the Care Assist mobile application (version 4.1.1 and earlier) would not play custom ringtones when iOS device screens were locked with a passcode.
-
Class IIofficial
Covidien, LP recalls Covidien Auto Suture" Structural Balloon TrocarThe potential for trocar seal disengagement when using mesh products incorrectly with the device.
-
Class IIofficial
Covidien, LP recalls Covidien Auto Suture" Blunt Tip TrocarThe potential for trocar seal disengagement when using mesh products incorrectly with the device.
-
Class IIofficial
Covidien, LP recalls Covidien Auto Suture" Blunt Tip TrocarThe potential for trocar seal disengagement when using mesh products incorrectly with the device.
-
Class IIofficial
Covidien, LP recalls Covidien Auto Suture" Blunt Tip TrocarThe potential for trocar seal disengagement when using mesh products incorrectly with the device.
-
Class IIofficial
Covidien, LP recalls Covidien Auto Suture" Blunt Tip TrocarThe potential for trocar seal disengagement when using mesh products incorrectly with the device.
-
Class IIofficial
Covidien, LP recalls Covidien Auto Suture" Blunt Tip TrocarThe potential for trocar seal disengagement when using mesh products incorrectly with the device.
-
Class IIofficial
Covidien, LP recalls Covidien Auto Suture" Blunt Tip TrocarThe potential for trocar seal disengagement when using mesh products incorrectly with the device.
-
Class IIofficial
Seaspine Orthopedics Corporation recalls Mariner RDX MIS Polyaxial HeadScrew heads, of pedicle screw systems, have partial thread within the screw head that pose a risk of intraoperative cross-threading within the locking cap/screw head interface.
-
Class Iofficial
T Marzetti Co recalls Chick-fil-a Polynesian Sauce Cup: 28g single serve food service cupUndeclared Allergens - Wheat and Soy
-
Class IIofficial
The Coca-cola Company recalls Powerade Zero Mixed Berry - 20 oz PET bottles 24 variety packThe potential presence of a foreign metal object in the product, i.e. a stainless-steel ring.
-
Class IIofficial
The Coca-cola Company recalls Powerade Zero Fruit Punch - 20oz PET bottles 24 variety packThe potential presence of a foreign metal object in the product, i.e. a stainless-steel ring.
-
Class IIofficial
The Coca-cola Company recalls Powerade Mountain Berry Blast - 20oz PET bottles 24 variety packThe potential presence of a foreign metal object in the product, i.e. a stainless-steel ring.
-
Class Iofficial
Roland Foods, LLC. recalls Roland Tahini; 100% Ground Sesame Seeds; Net WT. 16 OZ. (1 LB.) 454gProduct may be contaminated with Salmonella.
-
Class IIofficial
Ortho-Clinical Diagnostics recalls VITROS Performance Verifier IThe issue was mismatched diluent lots assembled into VITRO Chemistry Products Performance Verifier I and II. This issue has the potential to cause erroneous Quality Control (QC) results that could lead to testing delays and erroneous patient results. This issue caused positive or negative bias in the quantification of sodium (Na+) and total carbon dioxide (ECO2). The incorrect diluent into the Performance Verifier II exhibited a potential negative bias of up to 6 mmol/L for ECO2 and Na+ testing.
-
Class IIofficial
Ortho-Clinical Diagnostics recalls VITROS Performance Verifier IIThe issue was mismatched diluent lots assembled into VITRO Chemistry Products Performance Verifier I and II. This issue has the potential to cause erroneous Quality Control (QC) results that could lead to testing delays and erroneous patient results. This issue caused positive or negative bias in the quantification of sodium (Na+) and total carbon dioxide (ECO2). The incorrect diluent into the Performance Verifier II exhibited a potential negative bias of up to 6 mmol/L for ECO2 and Na+ testing.
-
Class IIofficial
Emergency Products & Research recalls O.T.D - Optimum Traction Device - Street Orange. Single use traction splint.Failure of adhesive to secure the pole insert.