Class Iofficial
Wawa Beverage Company recalls Wawa Fruit Punch Pint (16oz)Undeclared Milk Allergen,
-
Class Iofficial
Cook Incorporated recalls Centimeter Sizing Catheter RPN N5.0-35-100-P-10S-PIG-CSC-20 GPN G11916 Catheters for use in angiographic proceMarker bands may be at an increased risk of cracking/breakage. Potential adverse events that may result include increased procedural time, harms associated with device fragmentation/separation, and vessel injury. In a worst-case scenario, device fragmentation and separation may cause life-threatening harm or death.
-
Class Iofficial
Cook Incorporated recalls Aurous Centimeter Sizing Catheter RPN N5.0-35-100-P-10S-PIG-CSC-20-01 GPN G47308 Sizing catheters are single-uMarker bands may be at an increased risk of cracking/breakage. Potential adverse events that may result include increased procedural time, harms associated with device fragmentation/separation, and vessel injury. In a worst-case scenario, device fragmentation and separation may cause life-threatening harm or death.
-
Class Iofficial
Cook Incorporated recalls Beacon Tip Centimeter Sizing Catheter RPN/GPN NR5.0-35-70-P-10S-0-CSC-20 G31213 NR5.0-35-100-P-10S-0-CSC-20 G3Marker bands may be at an increased risk of cracking/breakage. Potential adverse events that may result include increased procedural time, harms associated with device fragmentation/separation, and vessel injury. In a worst-case scenario, device fragmentation and separation may cause life-threatening harm or death.
-
Class IIofficial
Merit Medical Systems, Inc. recalls Philips Bridge Prep Kit REF: K12-09098B Sterile EOcatheter may experience resistance when being advanced over the guidewire
-
Class IIofficial
Merit Medical Systems, Inc. recalls Philips Bridge Prep Kit REF: K12-09098C Sterile EOcatheter may experience resistance when being advanced over the guidewire
-
Class IIofficial
Merit Medical Systems, Inc. recalls Philips Bridge Prep Kit REF: K12-09098C Sterile EOcatheter may experience resistance when being advanced over the guidewire
-
Class IIofficial
Siemens Healthcare Diagnostics Inc recalls epoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System; Siemens Material Number (SMN): 10736515;Use of the affected product can result in discrepant low pH and measured Total Carbon Dioxide (mTCO2) results and corresponding calculated values in arterial, venous, and capillary patient samples. An erroneously low pH or mTCO2 reading can lead to inappropriate treatment of nonexistent acidosis (respiratory of metabolic) or missed diagnosis of alkalosis (a true high pH: respiratory or metabolic alkalosis), potentially causing harm through unnecessary interventions with the former or failure to deliver appropriate treatment with the latter.
-
Class IIofficial
Intuitive Surgical, Inc. recalls Brand Name: Da Vinci X and Da Vinci Xi Product Name: da Vinci XDue to manipulator containing screws with in the arm sub-assembly that my be susceptible to breaking.
-
Class Iofficial
Windstone Medical Packaging, Inc. recalls Custom Convenience Kit containing recalled Medline Namic Angiographic Control Syringes with Rotating Adapter (Affected kits contain Medline Namic RA Syringes, which are recalled by their manufacturer due to risk of syringe rotating adapter unwinding during use, which may result in a loose connection or full disconnection between syringe and manifold. If unwinding occurs, there is a potential for biohazard exposure, blood loss, infection. or air embolism.
-
Class IIofficial
DFI Co., Ltd. recalls One Step 10A in vitro diagnostic testThe devices were distributed without required FDA premarket clearance or approval.
-
Class IIofficial
DFI Co., Ltd. recalls One Step UTI in vitro diagnostic test REF: 3374The devices were distributed without required FDA premarket clearance or approval.
-
Class IIofficial
DFI Co., Ltd. recalls One Step pH in vitro diagnostic test REF: 31I4PThe devices were distributed without required FDA premarket clearance or approval.
-
Class IIofficial
DFI Co., Ltd. recalls QUCARE Total Cholesterol in vitro diagnostic test REF: 6407The devices were distributed without required FDA premarket clearance or approval.
-
Class IIofficial
DFI Co., Ltd. recalls Uric Acid in vitro diagnostic test REF: 31H0PThe devices were distributed without required FDA premarket clearance or approval.
-
Class IIofficial
DFI Co., Ltd. recalls One Step K in vitro diagnostic test REF: 81A4The devices were distributed without required FDA premarket clearance or approval.
-
Class IIofficial
DFI Co., Ltd. recalls One Step P in vitro diagnostic test REF: 8194The devices were distributed without required FDA premarket clearance or approval.
-
Class IIofficial
International Medication Systems Ltd. recalls EPINEPHRINE InjectionLack of Assurance of Sterility
-
Class IIofficial
Island Kinetics, Inc. d.b.a. CoValence Laboratories recalls Blemish Spot Treatment (Salicylic Acid) 1%Failed Stability Specifications
-
Class IIofficial
Wells Pharma of Houston LLC recalls Phenylephrine Hydrochloride Injectable SolutioncGMP deviations.
-
Class IIofficial
Wells Pharma of Houston LLC recalls fentaNYL Citrate Injectable SolutioncGMP deviations.
-
Class IIofficial
Wells Pharma of Houston LLC recalls fentaNYL Citrate injectable Solution in 0.9% Sodium ChloridecGMP deviations.
-
Class IIofficial
Wells Pharma of Houston LLC recalls fentaNYL Citrate Injectable SolutioncGMP deviations.
-
Class IIofficial
Wells Pharma of Houston LLC recalls fentaNYL Citrate Injectable Solution in 0.9% Sodium ChloridecGMP deviations.
-
Class IIofficial
Wells Pharma of Houston LLC recalls Ketamine Hydrochloride Injectable SolutioncGMP deviations.
-
Class IIofficial
Wells Pharma of Houston LLC recalls fentaNYL Citrate Injectable Solution in 0.9% Sodium ChloridecGMP deviations.
-
Class IIofficial
Pure Palm C/o Dvd recalls Label is predominantly green with white lettering in a clear plastic clamshell container. Pure Palm dates- Medmislabeled Coconut Date Bites. (product is coconut covered, label is for regular date bites)
-
Class IIofficial
Abbott Point Of Care Inc. recalls i-STAT EG7+ cartridge; List Number: 03P76-25;Approximately 7.6% of cartridges from specific i-STAT EG7+ (list number 03P76-25), i-STAT EG6+ (list number 03P77-25) and i-STAT G3+ (list number 03P78-26) cartridge lots in the U.S. marketplace may report higher than expected PCO2 and lower than expected pH results due to a manufacturing issue. A falsely high pCO2 together with a falsely low pH makes the blood gas appear as if the patient is in severe respiratory acidosis. If the clinician believes the values are real, it may trigger unnecessary or even harmful interventions such as unnecessary ventilator changes, unnecessary medications, potential harm from overcorrection, or delays in identifying true problems.
-
Class IIofficial
Abbott Point Of Care Inc. recalls i-STAT EG6+ cartridge; List Number: 03P77-25;Approximately 7.6% of cartridges from specific i-STAT EG7+ (list number 03P76-25), i-STAT EG6+ (list number 03P77-25) and i-STAT G3+ (list number 03P78-26) cartridge lots in the U.S. marketplace may report higher than expected PCO2 and lower than expected pH results due to a manufacturing issue. A falsely high pCO2 together with a falsely low pH makes the blood gas appear as if the patient is in severe respiratory acidosis. If the clinician believes the values are real, it may trigger unnecessary or even harmful interventions such as unnecessary ventilator changes, unnecessary medications, potential harm from overcorrection, or delays in identifying true problems.
-
Class IIofficial
Abbott Point Of Care Inc. recalls i-STAT G3+ cartridge; List Number: 03P78-26;Approximately 7.6% of cartridges from specific i-STAT EG7+ (list number 03P76-25), i-STAT EG6+ (list number 03P77-25) and i-STAT G3+ (list number 03P78-26) cartridge lots in the U.S. marketplace may report higher than expected PCO2 and lower than expected pH results due to a manufacturing issue. A falsely high pCO2 together with a falsely low pH makes the blood gas appear as if the patient is in severe respiratory acidosis. If the clinician believes the values are real, it may trigger unnecessary or even harmful interventions such as unnecessary ventilator changes, unnecessary medications, potential harm from overcorrection, or delays in identifying true problems.
-
Class IIofficial
BioFire Diagnostics, LLC recalls Biofire Spotfire Respiratory/Sore Throat Panel REF: 423485Respiratory/sore throat panel test may result in false negative results and control failures.
-
Class IIofficial
Aniara Diagnostica LLC recalls ZYMUTEST HIA IgGKits from lot FD1265 had been packaged on 17 October 2024 and put on the market with an outdated flyer (8 October 2024), containing incorrect values for the positive control.
-
Class IIIofficial
Avantor Performance Materials LLC recalls Magnesium ChlorideSubpotent drug
-
Class IIofficial
Erbe Medical, LLC recalls ERBEFLO CleverCap CO2: Hybrid CO2 Tubing/Cap Set for Olympus Scopes & CO2 SourcesDevices providing sterile water/CO2/air to endoscopes were manufactured with a distal irrigation segment connector configuration that may be more susceptible to unintended water flow from distal tip under unsupported third-party connector use conditions, which may lead to water aspiration, which may progress to aspiration pneumonia, acute respiratory distress syndrome, or respiratory failure.
-
Class IIofficial
Erbe Medical, LLC recalls ERBEFLO CleverCap: Hybrid Tubing/Cap Set for Olympus ScopesDevices providing sterile water/CO2/air to endoscopes were manufactured with a distal irrigation segment connector configuration that may be more susceptible to unintended water flow from distal tip under unsupported third-party connector use conditions, which may lead to water aspiration, which may progress to aspiration pneumonia, acute respiratory distress syndrome, or respiratory failure.
-
Class IIofficial
Erbe Medical, LLC recalls ERBEFLO 2Devices providing sterile water/CO2/air to endoscopes were manufactured with a distal irrigation segment connector configuration that may be more susceptible to unintended water flow from distal tip under unsupported third-party connector use conditions, which may lead to water aspiration, which may progress to aspiration pneumonia, acute respiratory distress syndrome, or respiratory failure.
-
Class IIofficial
Philips Medical Systems Nederland B.v. recalls Philips Azurion. System Code Description (Model Number): Azurion 3 M12 (722221Potential for the bolts and plastic parts of the Cable Hose Carriers that hold the cable hose to the monitor ceiling suspension may loosen and/or break over time due to the forces applied when the monitor is moved or rotated.
-
Class IIofficial
Heytea USA Inc recalls Coconut Drink 1; NET VOLUME: 1L; Ingredients: Coconut waterProduct contains undeclared milk.
-
Class IIofficial
Zimmer, Inc. recalls Brand Name: Disposable Mixing Bowls with Spatula Product Name: Mixing Bowl and Spatula Model/Catalog Number: 0Ten complaints have been received identifying issues at the time of use related to the package seal, including incomplete seals, wrinkles in the seals or peeling seals. Compromised sterile barrier may lead to clinically insignificant extension of surgery or infection leading to surgical intervention.
-
Class IIofficial
Boston Scientific Corporation recalls LUX-DX IIFor some patients upgraded from a LUX-Dx M301 device to a new Model M302 or M312, the new ICM device is not collecting PVC Burden data (both M302/M312) or monitoring for Bradyarrhythmia and Pause (M312 only). In addition, for these devices that are not being monitored, the LATITUDE Clarity programming screens have discrepant information, indicating that the monitoring is enabled for these features in one location and not enabled in another location.
-
Class IIofficial
Boston Scientific Corporation recalls LUX-Dx II PlusFor some patients upgraded from a LUX-Dx M301 device to a new Model M302 or M312, the new ICM device is not collecting PVC Burden data (both M302/M312) or monitoring for Bradyarrhythmia and Pause (M312 only). In addition, for these devices that are not being monitored, the LATITUDE Clarity programming screens have discrepant information, indicating that the monitoring is enabled for these features in one location and not enabled in another location.
-
Class IIofficial
Appco Pharma LLC recalls Prazosin Hydrochloride CapsulescGMP deviation: detection of Nitrosamine Drug Substance-Related Impurities, N-nitroso-prazosin impurity C (NNP) above acceptable limits.
-
Class IIofficial
Appco Pharma LLC recalls Prazosin Hydrochloride CapsulescGMP deviation: detection of Nitrosamine Drug Substance-Related Impurities, N-nitroso-prazosin impurity C (NNP) above acceptable limits.
-
Class IIofficial
Appco Pharma LLC recalls Prazosin Hydrochloride CapsulescGMP deviation: detection of Nitrosamine Drug Substance-Related Impurities, N-nitroso-prazosin impurity C (NNP) above acceptable limits.
-
Class IIofficial
Blueroot Health, Inc. recalls Vital Nutrients Aller-C (vitamin CContains undeclared egg, soy, and hazelnut
-
Class IIofficial
Micro-X Ltd. recalls Rover Mobile X-ray System Model/Catalog Number: MXU-RV35 and MXU-RV71X-Ray generator may malfunction resulting in the X-Ray being inoperable.
-
Class Iofficial
North American Rescue LLC. recalls AIDBAGs are first aid kits of convenience composed of individually labeled devices designed as specialty firstKits contain the McKesson TRUE Metrix PRO Professional Monitoring Blood Glucose System that has been recalled for an issue with the software design of the E-5 Error Code where the meter displays an E-5 error code for a very high blood glucose event (> 600 mg/dL) or when there is a test strip error, which can lead to operator confusion when evaluating patients and a delay in appropriate treatment. This is a downstream recall of Trividia Health Class I, Recall Event ID 98317.
-
Class IIofficial
Straumann USA LLC recalls Straumann n!ce PMMA Full-arch Restoration. Screw-retained Bridge. Article 010.0304. Endosseous Dental ImplantIncludes an incorrect screw seat interface.
-
Class IIofficial
Straumann USA LLC recalls Straumann n!ce ZrIncludes an incorrect screw seat interface.
-
Class IIofficial
Edwards Lifesciences, LLC recalls Edwards EVOQUE tricuspid delivery system REF:9850TDS Sterile EOLabeling update to provide warning if functionality of the valve replacement delivery system is compromised.