Noah Medical recalls
Noah Medical is named on 5 recalls in Recall Relay's records since November 2023, from FDA; the most recent was September 12, 2025.
All 5, newest first.
-
Class IIofficial
Noah Medical Corporation recalls Brand Name: Galaxy System Product Name: Galaxy System Model/Catalog Number: GAL-001 Product Description: The GDue to Users not having access to the Instructions for Use/User Manual due to it being password protected.
-
Class IIofficial
Noah Medical recalls Galaxy System (GAL-001). Intended to provide bronchoscopic visualization of and access to patient airways forBronchoscopic visualization system power supplies have defective semiconductor components that may fail unexpectedly, which could result in an aborted procedure since it is a non recoverable failure. Manual removal of the bronchoscope when the system is not functioning can increase the chance of injury to the tissues of the lung as the scope is withdrawn including the possibility of pneumothorax.
-
Class IIofficial
Noah Medical recalls Galaxy System (GAL-001)Bronchoscopic system uninterruptable power may have defective component, may cause power supply short circuit/trip protective circuit breakers causing sudden, unrecoverable power loss, which could occur while system is on mains or internal battery power; if during procedure, could cause interrupted/aborted procedure, may necessitate manual removal, which could lead to lung injury/pneumothorax.
-
Class IIofficial
Noah Medical recalls Galaxy bronchoscope (Product GALB-001) used with Instructions for useNot all biopsy tools meant to fit bronchoscope channel (2.1mm inner diameter/length greater or equal to 1100mm) will pass through if scope channel positioned in sharp bend (approx. greater than 100-deg.), excessive force may result in tissue injury/pneumothorax; instruments with long rigid distal sections may exhibit insertion resistance when bronchoscope bending section significantly articulated.
-
Class IIofficial
Noah Medical recalls Galaxy System REF GAL-001 (specifically GAL-019Due to a misaligned calibration of the wired controller that may result in unintended motion of the scope.
This page groups recalls by the company name the agency printed, so it can include different companies with similar names, and a recall may name a store or distributor rather than the maker. It counts the recalls in Recall Relay's records, not a company's whole history; an FDA recall that lists several products counts once. Open a recall to see its details and the agency's official notice, which is always the final word. Recall Relay isn't affiliated with any agency or with the companies named here.
Severity labels say where they come from: “official” is the agency's own class, “advisory” is based on the agency's safety warning, and “est. risk” is Recall Relay's estimate from the hazard description, not an agency classification.
Hear about the next one
Recall Relay can email you when a new recall mentions Noah Medical or any words you choose. Start with a free key (no card needed), then set up alerts on your account page.
Get recall alerts