Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Noah Medical recalls Galaxy System REF GAL-001 (specifically GAL-019

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Due to a misaligned calibration of the wired controller that may result in unintended motion of the scope.
Products
  • Galaxy System REF GAL-001 (specifically GAL-019
Where it was sold

U.S. Nationwide distribution in the states of CA, IN, MD, MO, OH, and PA.

States: California, Indiana, Maryland, Missouri, Ohio, Pennsylvania

Barcodes
00850048825048
Units affected
17 wired controllers
About this recall
Galaxy System REF GAL-001 (specifically GAL-019, Wired Controller) The Galaxy System and its accessories are intended to provide bronchoscopic visualization of and access to patient airways for diagnostic and therapeutic procedures.