Noah Medical recalls Galaxy System (GAL-001)
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Bronchoscopic system uninterruptable power may have defective component, may cause power supply short circuit/trip protective circuit breakers causing sudden, unrecoverable power loss, which could occur while system is on mains or internal battery power; if during procedure, could cause interrupted/aborted procedure, may necessitate manual removal, which could lead to lung injury/pneumothorax.
- Products
-
- Galaxy System (GAL-001)
- Company
- Noah Medical
- Where it was sold
-
US: CA, MA, PA, NJ, ID, CT, MO, NC, MD, IL, TN, VA, OH, IN, WV, NY
States: California, Connecticut, Idaho, Illinois, Indiana, Massachusetts, Maryland, Missouri, North Carolina, New Jersey, New York, Ohio, Pennsylvania, Tennessee, Virginia, West Virginia
- Barcodes
00850048825048- Units affected
- 26
- About this recall
- Galaxy System (GAL-001)