Noah Medical recalls Galaxy System (GAL-001). Intended to provide bronchoscopic visualization of and access to patient airways for

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Bronchoscopic visualization system power supplies have defective semiconductor components that may fail unexpectedly, which could result in an aborted procedure since it is a non recoverable failure. Manual removal of the bronchoscope when the system is not functioning can increase the chance of injury to the tissues of the lung as the scope is withdrawn including the possibility of pneumothorax.
Products
  • Galaxy System (GAL-001). Intended to provide bronchoscopic visualization of and access to patient airways for
Where it was sold

US: SC, OH, PA, TN, MO, IN, CA, WV, NY, MD, MA, NJ, ID, CT, NC

States: California, Connecticut, Idaho, Indiana, Massachusetts, Maryland, Missouri, North Carolina, New Jersey, New York, Ohio, Pennsylvania, South Carolina, Tennessee, West Virginia

Barcodes
00850048825048
Units affected
18
About this recall
Galaxy System (GAL-001). Intended to provide bronchoscopic visualization of and access to patient airways for diagnostic and therapeutic procedures.