GE Medical Systems, LLC recalls MAC VU360 Acquisition Trunk Cable and Module Holder
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- If a user incorrectly places the Acquisition Module into the Acquisition Module Holder, it can result in excessive bending and wear of the Acquisition Trunk Cable. This is visible as a damaged sheath or exposed wires. If a damaged Acquisition Trunk Cable is not recognized and removed from service, the cable can overheat with continued use.
- Products
-
- MAC VU360 Acquisition Trunk Cable and Module Holder
- Company
- GE Medical Systems, LLC
- Where it was sold
-
Worldwide Distribution: US (nationwide) and OUS (foreign/Global) to countries of: Australia, Austria, Bahrain, Belgium, Canada, Czech Republic, Denmark, Finland, France, Germany, Hong Kong, Ireland, Israel, Italy, Japan, Korea, Republic of, Kuwait, Malaysia, Netherlands, New Zealand, Norway, Philippines, Puerto Rico, Qatar, Saudi Arabia, Singapore, Sweden, Switzerland, Thailand, United Arab Emirates, and United Kingdom.
Nationwide
- Barcodes
00840682125499- Units affected
- 29,225 units
- About this recall
- MAC VU360 Acquisition Trunk Cable and Module Holder, high performance, multichannel resting electrocardiograph, Model Number 2030360-001, Product Codes VU2 and VU4