Leica Biosystems Nussloch Gmbh recalls Leica HistoCore PEGASUS (PROCESSOR
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- There is an issue regarding poorly processed and/or damaged biopsy tissue specimens on the HistoCore PEGASUS / HistoCore PEGASUS Plus resulting from an incorrect assignment of the carryover value set for created or edited protocols.
- Products
-
- Leica HistoCore PEGASUS (PROCESSOR
- Where it was sold
-
US, Nationwide
Nationwide
- Units affected
- 98 devices
- About this recall
- Leica HistoCore PEGASUS (PROCESSOR, TISSUE, AUTOMATED)