GE Medical Systems, LLC recalls GE Healthcare Optima XR220amx
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement performed, some bolts used in the Column Safety mechanism can be missing. If these bolts are missing and a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work properly causing the arm with X-ray tube and collimator to drop downward on the column track.
- Products
-
- GE Healthcare Optima XR220amx
- Company
- GE Medical Systems, LLC
- Where it was sold
-
Worldwide
Nationwide
- Barcodes
840682115261- Units affected
- 281 units
- About this recall
- GE Healthcare Optima XR220amx, Mobile Digital Radiographic System, Model Numbers: 1) 5555000-12; 2) 5555000-5; 3) 5555000-6; 4) A0659PB; 5) to be provided