Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

GE Medical Systems, LLC recalls GE Healthcare Optima XR240amx

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement performed, some bolts used in the Column Safety mechanism can be missing. If these bolts are missing and a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work properly causing the arm with X-ray tube and collimator to drop downward on the column track.
Products
  • GE Healthcare Optima XR240amx
Where it was sold

Worldwide

Nationwide

Barcodes
840682122351
Units affected
120 units
About this recall
GE Healthcare Optima XR240amx, Mobile Digital Radiographic System, Model Numbers: 1) 5555000-27; 2) 5555000-31; 3) 5555000-33; 4) 5555000-41; 5) 5555000-47; 6) A0700DL; 7) A1700DA; 8) To be provided.