Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

GE Medical Systems, LLC recalls Tec 6 Plus Anesthesia Vaporizer

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Due to a manufacturing issue, certain Tec 6 Plus and Tec 800 series vaporizers can deliver less anesthetic agent than the vaporizer dial setting.
Products
  • Tec 6 Plus Anesthesia Vaporizer
Where it was sold

US Nationwide. Global Distribution.

Nationwide

Barcodes
008406821126590084068211257400840682112925008406821128640084068211323600840682113069008406821126660084068211285700840682112505008406821125120084068211243700840682112802008406821127100084068211276500840682112789008406821128260084068211590200840682112451
Units affected
1573 units
About this recall
Tec 6 Plus Anesthesia Vaporizer, Reference Numbers 1107-9601-000, 1107-9602-000, 1107-9603-000, 1107-9604-000, 1107-9610-000, 1107-9612-000, 1107-9616-000, 1107-9617-000, 2066835-001, 2069767-001, 2078800-001, 2091008-001, 2091009-001, M1098876, M1145610, M1220023, M1220854, M1221495, and M1233992