GE Medical Systems, LLC recalls Portrait Core Services
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- There is a software issue that can result in the loss of patient monitoring on the Portrait Central Viewer Application after 425 days of continuous run-time.
- Products
-
- Portrait Core Services
- Company
- GE Medical Systems, LLC
- Where it was sold
-
Worldwide - US Nationwide distribution in the states of OH, TX and the countries of Australia, China, Czech Republic, France, Italy, Netherlands, Qatar, Singapore, UAE, UK.
States: Ohio, Texas
- Barcodes
00195278659446- Units affected
- 7 systems
- About this recall
- Portrait Core Services, a set of software services that enable the communication and interaction of the Portrait Mobile Monitoring Solution components and will integrate into existing healthcare facility infrastructure and clinical information systems, Model CSSXB