GE Medical Systems, LLC recalls Optima XR646 HD. X-Ray imaging system.
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- The user can continue to take the next exposure after reaching the Automatic Exposure Control (AEC) limit, without taking any further actions to acknowledge or adjust exposure techniques. In all cases, a message still appears on the screen when the AEC limit is reached.
- Products
-
- Optima XR646 HD. X-Ray imaging system.
- Company
- GE Medical Systems, LLC
- Where it was sold
-
Worldwide Distribution: US (nationwide) including Puerto Rico and OUS (foreign) countries to: Canada, Chile, China, Colombia, Coast Rica, Croatia, Cyprus, Czech Republic, Czechia, Dominican Republic, Egypt, Estonia, Finland, France, Germany, Gibraltar, Greece, Honduras, Hong Kong, Hungary, Iceland, Indonesia, Iraq, Ireland, Italy, Jamaica, Japan, Jordan, Kenya, Korea (Republic of) Kuwait, Libya, Lithuania, Macedonia, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Nigeria, Norway, Oman, Pakistan, Panama, Paraguay, Philippines, Poland, Portugal, Qatar, Republic of Serbia, Romania, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, and Vietnam.
Nationwide
- Barcodes
00195278526892- Units affected
- 3149 (1697 US; 1452 OUS)
- About this recall
- Optima XR646 HD. X-Ray imaging system.