GE Medical Systems, LLC recalls GE Healthcare Revolution Apex Select

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
GE HealthCare has become aware of a potential issue on certain Revolution CT and Revolution Apex series CT systems (please see affected product details and the Appendix) that can result in coolant fluid (glycol) leaking onto the floor near the base of the gantry. The fluid is visible as it has a blue color. It is not corrosive and is not at a harmful temperature. However, there is a potential for injury if the fluid is not noticed and a person slips and/or falls.
Products
  • GE Healthcare Revolution Apex Select
Where it was sold

Worldwide distribution.

Nationwide

Barcodes
00195278460271
Units affected
1 unit
About this recall
GE Healthcare Revolution Apex Select, System, X-ray, Tomography, Computed