GE Medical Systems, LLC recalls GE Healthcare Revolution CT ES
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- GE HealthCare has become aware of a potential issue on certain Revolution CT and Revolution Apex series CT systems (please see affected product details and the Appendix) that can result in coolant fluid (glycol) leaking onto the floor near the base of the gantry. The fluid is visible as it has a blue color. It is not corrosive and is not at a harmful temperature. However, there is a potential for injury if the fluid is not noticed and a person slips and/or falls.
- Products
-
- GE Healthcare Revolution CT ES
- Company
- GE Medical Systems, LLC
- Where it was sold
-
Worldwide distribution.
Nationwide
- Barcodes
0019527846027100195278633149008406821466160084068212347100840682118552- Units affected
- 21 units
- About this recall
- GE Healthcare Revolution CT ES, System, X-ray, Tomography, Computed