Leica Biosystems Nussloch Gmbh recalls Leica HistoCore PEGASUS Tissue Processor
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Poorly processed and/or damaged biopsy tissue specimens on the device resulting from reagent levels exceeding the maximum fill level marks on reagent bottles or in the paraffin tanks.
- Products
-
- Leica HistoCore PEGASUS Tissue Processor
- Where it was sold
-
Worldwide - US Nationwide distribution including in the states of AK, AL, AZ, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MT, NC, NJ, NV, NY, OH, TN, TX, UT, VA, WA, and WI. There was also government distribution but no military distribution. The country of Canada.
Nationwide
- Barcodes
04049188216397- Units affected
- 141 devices
- About this recall
- Leica HistoCore PEGASUS Tissue Processor, REF 14048858005.