Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

GE Medical Systems, LLC recalls SIGNA Architect

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
GE HealthCare has become aware that gradient coils for certain MR systems (see affected product list in this letter), under specific conditions, can produce elevated acoustic noise during scanning. As a result, acoustic levels can exceed 99dB, the limit established by International Electrotechnical Commission (IEC 60601-2-33) when using hearing protection with a Noise Reduction Rating (NRR) of 29dB, as currently required in the Operator Manual.
Products
  • SIGNA Architect
Where it was sold

Worldwide distribution - US Nationwide and the country of China.

Nationwide

Barcodes
00840682147095
Units affected
N/A
About this recall
SIGNA Architect