GE Medical Systems, LLC recalls SIGNA PET/MR
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- GE HealthCare has become aware that gradient coils for certain MR systems (see affected product list in this letter), under specific conditions, can produce elevated acoustic noise during scanning. As a result, acoustic levels can exceed 99dB, the limit established by International Electrotechnical Commission (IEC 60601-2-33) when using hearing protection with a Noise Reduction Rating (NRR) of 29dB, as currently required in the Operator Manual.
- Products
-
- SIGNA PET/MR
- Company
- GE Medical Systems, LLC
- Where it was sold
-
Worldwide distribution - US Nationwide and the country of China.
Nationwide
- Barcodes
00840682105378- Units affected
- N/A
- About this recall
- SIGNA PET/MR