Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

GE Medical Systems, LLC recalls Allia IGS 5 Pulse angiographic X-ray system

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
GE HealthCare has become aware of a potential issue affecting an electrical cable on certain IGS systems equipped with Gaia generators. This issue could lead to a risk of electrical shock to service personnel performing installation, maintenance, or troubleshooting activities inside the C-FRT cabinet while power is ON.
Products
  • Allia IGS 5 Pulse angiographic X-ray system
Where it was sold

Worldwide - US Nationwide distribution.

Nationwide

Barcodes
00195278719263
Units affected
121 units
About this recall
Allia IGS 5 Pulse angiographic X-ray system