GE Medical Systems, LLC recalls GE HealthCare OEC 9900 Elite C-arm Systems.
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- GE HealthCare has become aware of a non-conformance to 21 CFR 1020.30 (k) in certain OEC 9800 and OEC 9900 Elite C-arm Systems. Specifically, impacted systems are missing the full layer of lead tape in the collimator covers, which will lead to a small increase in radiation exposure to the patient and/or operator. The missing lead tape is not detectable by the user.
- Products
-
- GE HealthCare OEC 9900 Elite C-arm Systems.
- Company
- GE Medical Systems, LLC
- Where it was sold
-
US Nationwide distribution.
Nationwide
- Barcodes
00840682114349- About this recall
- GE HealthCare OEC 9900 Elite C-arm Systems.