GE Medical Systems, LLC recalls SIGNA PET/MR
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
- Products
-
- SIGNA PET/MR
- Company
- GE Medical Systems, LLC
- Where it was sold
-
Worldwide distribution.
Nationwide
- Units affected
- 123 units
- About this recall
- SIGNA PET/MR, with affected software versions: MP24.0, MP26.0, MP26.1, MR30.0, MR30.1; Tomographic imager combining emission computed tomography with nuclear magnetic resonance