FDAMedical devicesRecalled Status: Ongoing

GE Medical Systems, LLC recalls Optima MR360

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
Products
  • Optima MR360
Where it was sold

Worldwide distribution.

Nationwide

Units affected
469 units
About this recall
Optima MR360, with affected software versions: SV20.3, SV23.3, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System