Leica Biosystems Nussloch Gmbh recalls Leica HistoCore Arcadia H

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
An issue with safe usage of the device was identified whereby toxic smoke and internal fire developed resulting from a problem associated with the sealing of the paraffin tank.
Products
  • Leica HistoCore Arcadia H
Where it was sold

US Nationwide, including government distribution.

Nationwide

Barcodes
0404918819476304049188205971
Units affected
180 devices
About this recall
Leica HistoCore Arcadia H, REF: 14039354100 and 14039357258, Paraffin Embedding Station.