Leica Biosystems Nussloch Gmbh recalls Leica HistoCore Arcadia H
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- An issue with safe usage of the device was identified whereby toxic smoke and internal fire developed resulting from a problem associated with the sealing of the paraffin tank.
- Products
-
- Leica HistoCore Arcadia H
- Where it was sold
-
US Nationwide, including government distribution.
Nationwide
- Barcodes
0404918819476304049188205971- Units affected
- 180 devices
- About this recall
- Leica HistoCore Arcadia H, REF: 14039354100 and 14039357258, Paraffin Embedding Station.