Class IIofficial
Beckman Coulter, Inc. recalls DxU 840m IrisThe rinse-line check valve has a shorter service lifespan than originally expected and requires annual replacement to prevent lamina fluid leakage; however, the check valve replacement was excluded from the annual preventive maintenance (PM) requirements without adequate reliability evaluation. When the check valve is not replaced annually, it may allow lamina fluid to migrate internally and leak onto the table or floor, potentially resulting in exposure to physical hazards (slip/fall), chemical hazards (lamina fluid exposure), biohazardous exposure, property damage, and delayed patient results.
-
Class IIofficial
Beckman Coulter, Inc. recalls IQ200 SPRINTThe rinse-line check valve has a shorter service lifespan than originally expected and requires annual replacement to prevent lamina fluid leakage; however, the check valve replacement was excluded from the annual preventive maintenance (PM) requirements without adequate reliability evaluation. When the check valve is not replaced annually, it may allow lamina fluid to migrate internally and leak onto the table or floor, potentially resulting in exposure to physical hazards (slip/fall), chemical hazards (lamina fluid exposure), biohazardous exposure, property damage, and delayed patient results.
-
Class Iofficial
Abiomed, Inc. recalls Automated Impella Controller (AIC) with the below product descriptions and corresponding Product Codes. 1. ImpThe potential for purge cassette recognition issues due to the failure of the purge flag component of the purge pressure sensor assembly.
-
Class Iofficial
Abiomed, Inc. recalls Impella Optical Controller with the below product descriptions and corresponding Product Codes. 1. Impella OptThe potential for purge cassette recognition issues due to the failure of the purge flag component of the purge pressure sensor assembly.
-
Class Iofficial
Abiomed, Inc. recalls Impella Optical ControllerThe potential for purge cassette recognition issues due to the failure of the purge flag component of the purge pressure sensor assembly.
-
Class IIofficial
Straumann USA LLC recalls GM Helix ImplantThere is a discrepancy between the product identification on the packaging and the dimensions of the implant contained inside.
-
Class IIofficial
CooperVision, Inc. recalls Brand Name: Pearle Vision 1 day Product Name: (somofilcon A) Contact Lens Model/Catalog Number: Pearle VisionA small number of lenses within the affected lot are of the incorrect power.
-
Class IIofficial
Intervascular, Sas recalls Cardioroot Woven Collagen Coated Vascular Graft; Model Numbers: 1. HEWROOT0024Product failed to meet the specified acceptance criteria for Integral Water Permeability (IWP).
-
Class IIofficial
Karl Storz Endoscopy recalls HOPKINS Telescope 30¿Due to incorrect inside packaging label for eyepiece.
-
Class IIofficial
Baxter Healthcare Corporation recalls Baxter Novum IQ Syringe PumpA software issue that may occur during a Volume/Time (VOT) multi-syringe infusion when a syringe reaches empty and is replaced while a Volume To Be Infused (VTBI) remains.
-
Class IIofficial
Baxter Healthcare Corporation recalls Baxter Novum IQ Syringe PumpA software issue that may occur during a Volume/Time (VOT) multi-syringe infusion when a syringe reaches empty and is replaced while a Volume To Be Infused (VTBI) remains.
-
Class Iofficial
Medline Industries, LP recalls Medline Convenience Kits: 1) KIT ANESTHESIA L&D A- LINE SETThe kits contain certain lots of sterile applicators which may contain fungal contamination that may occur under certain environmental conditions, allowing the growth of Aspergillus penicillioides. Contamination of skin preparation products with Aspergillus penicillioides may lead to serious adverse health consequences, including systemic infection, sepsis, illness, and death.
-
Class Iofficial
Medline Industries, LP recalls Medline Convenience Kits: 1) OPEN HEART ADULTThe kits contain certain lots of sterile applicators which may contain fungal contamination that may occur under certain environmental conditions, allowing the growth of Aspergillus penicillioides. Contamination of skin preparation products with Aspergillus penicillioides may lead to serious adverse health consequences, including systemic infection, sepsis, illness, and death.
-
Class Iofficial
Medline Industries, LP recalls Medline Convenience Kits: 1) PORT ACCESS KIT UNIV OF ALAThe kits contain certain lots of sterile applicators which may contain fungal contamination that may occur under certain environmental conditions, allowing the growth of Aspergillus penicillioides. Contamination of skin preparation products with Aspergillus penicillioides may lead to serious adverse health consequences, including systemic infection, sepsis, illness, and death.
-
Class Iofficial
Medline Industries, LP recalls Medline Convenience Kits: 1) ACC010308AThe kits contain certain lots of sterile applicators which may contain fungal contamination that may occur under certain environmental conditions, allowing the growth of Aspergillus penicillioides. Contamination of skin preparation products with Aspergillus penicillioides may lead to serious adverse health consequences, including systemic infection, sepsis, illness, and death.
-
Class Iofficial
Medline Industries, LP recalls Medline Convenience Kits: 1) DYKM1000BThe kits contain certain lots of sterile applicators which may contain fungal contamination that may occur under certain environmental conditions, allowing the growth of Aspergillus penicillioides. Contamination of skin preparation products with Aspergillus penicillioides may lead to serious adverse health consequences, including systemic infection, sepsis, illness, and death.
-
Class Iofficial
Medline Industries, LP recalls Medline Convenience Kits: 1) DYNDH1193The kits contain certain lots of sterile applicators which may contain fungal contamination that may occur under certain environmental conditions, allowing the growth of Aspergillus penicillioides. Contamination of skin preparation products with Aspergillus penicillioides may lead to serious adverse health consequences, including systemic infection, sepsis, illness, and death.
-
Class Iofficial
Medline Industries, LP recalls Medline Convenience Kits: 1) DYNDB3086DThe kits contain certain lots of sterile applicators which may contain fungal contamination that may occur under certain environmental conditions, allowing the growth of Aspergillus penicillioides. Contamination of skin preparation products with Aspergillus penicillioides may lead to serious adverse health consequences, including systemic infection, sepsis, illness, and death.
-
Class Iofficial
CooperSurgical, Inc. recalls INCA (Infant Nasal Cannulae Assy) Infant Nasal CPAPPotential to exhibit a loose connection between the flexible tubing and the nasal cannula, during assembly or while in use, which may result in loss of CPAP pressure, possibly reducing the effectiveness of treatment and lead to decreased oxygenation until connection is restored.
-
Class Iofficial
CooperSurgical, Inc. recalls INCA (Infant Nasal Cannulae Assy) Neonatal Nasal CPAPPotential to exhibit a loose connection between the flexible tubing and the nasal cannula, during assembly or while in use, which may result in loss of CPAP pressure, possibly reducing the effectiveness of treatment and lead to decreased oxygenation until connection is restored.
-
Class IIofficial
Becton, Dickinson and Company recalls ForeSight Elite Tissue Oximetry Cable REF HEMFSM10 UDI-DI code: 00690103210927 The non-invasive ForeSight oximDue to incorrect switching of Channel 1 and 2 on the module and flag labels on the tissue oximetry cables.
-
Class IIofficial
Shanghai United Imaging Healthcare Co., Ltd. recalls Computed Tomography X-ray SystemPositron emission tomography and computed tomography x-ray systems have a software problem when Split or Local function is enabled, and multiple RIS exams are merged into a single study, that prevents patient images from being visible on the Patient Administration interface, which may result in the need for rescans.
-
Class IIofficial
Shanghai United Imaging Healthcare Co., Ltd. recalls Computed Tomography X-ray SystemPositron emission tomography and computed tomography x-ray systems have a software problem when Split or Local function is enabled, and multiple RIS exams are merged into a single study, that prevents patient images from being visible on the Patient Administration interface, which may result in the need for rescans.
-
Class IIofficial
Shanghai United Imaging Healthcare Co., Ltd. recalls Computed Tomography X-ray SystemPositron emission tomography and computed tomography x-ray systems have a software problem when Split or Local function is enabled, and multiple RIS exams are merged into a single study, that prevents patient images from being visible on the Patient Administration interface, which may result in the need for rescans.
-
Class IIofficial
Shanghai United Imaging Healthcare Co., Ltd. recalls Computed Tomography X-ray SystemPositron emission tomography and computed tomography x-ray systems have a software problem when Split or Local function is enabled, and multiple RIS exams are merged into a single study, that prevents patient images from being visible on the Patient Administration interface, which may result in the need for rescans.
-
Class IIofficial
Shanghai United Imaging Healthcare Co., Ltd. recalls Computed Tomography X-ray SystemPositron emission tomography and computed tomography x-ray systems have a software problem when Split or Local function is enabled, and multiple RIS exams are merged into a single study, that prevents patient images from being visible on the Patient Administration interface, which may result in the need for rescans.
-
Class IIofficial
Shanghai United Imaging Healthcare Co., Ltd. recalls Computed Tomography X-ray SystemPositron emission tomography and computed tomography x-ray systems have a software problem when Split or Local function is enabled, and multiple RIS exams are merged into a single study, that prevents patient images from being visible on the Patient Administration interface, which may result in the need for rescans.
-
Class IIofficial
Shanghai United Imaging Healthcare Co., Ltd. recalls Positron Emission Tomography and Computed Tomography SystemPositron emission tomography and computed tomography x-ray systems have a software problem when Split or Local function is enabled, and multiple RIS exams are merged into a single study, that prevents patient images from being visible on the Patient Administration interface, which may result in the need for rescans.
-
Class IIofficial
Shanghai United Imaging Healthcare Co., Ltd. recalls Positron Emission Tomography and Computed Tomography SystemPositron emission tomography and computed tomography x-ray systems have a software problem when Split or Local function is enabled, and multiple RIS exams are merged into a single study, that prevents patient images from being visible on the Patient Administration interface, which may result in the need for rescans.
-
Class IIofficial
Shanghai United Imaging Healthcare Co., Ltd. recalls Positron Emission Tomography and Computed Tomography SystemPositron emission tomography and computed tomography x-ray systems have a software problem when Split or Local function is enabled, and multiple RIS exams are merged into a single study, that prevents patient images from being visible on the Patient Administration interface, which may result in the need for rescans.
-
Class IIofficial
Shanghai United Imaging Healthcare Co., Ltd. recalls Positron Emission Tomography and Computed Tomography SystemPositron emission tomography and computed tomography x-ray systems have a software problem when Split or Local function is enabled, and multiple RIS exams are merged into a single study, that prevents patient images from being visible on the Patient Administration interface, which may result in the need for rescans.
-
Class IIofficial
Shanghai United Imaging Healthcare Co., Ltd. recalls Positron Emission Tomography and Computed Tomography SystemPositron emission tomography and computed tomography x-ray systems have a software problem when Split or Local function is enabled, and multiple RIS exams are merged into a single study, that prevents patient images from being visible on the Patient Administration interface, which may result in the need for rescans.
-
Class IIofficial
Shanghai United Imaging Healthcare Co., Ltd. recalls Positron Emission Tomography and Computed Tomography SystemPositron emission tomography and computed tomography x-ray systems have a software problem when Split or Local function is enabled, and multiple RIS exams are merged into a single study, that prevents patient images from being visible on the Patient Administration interface, which may result in the need for rescans.
-
Class IIofficial
AVID Medical, Inc. recalls DAVINCI TRUCUSTOM BUNDLE; Model/Item Number (Kit): 1. SAMM064-27Product may be potentially contaminated with a lubricant used during the manufacturing process.
-
Class IIofficial
AVID Medical, Inc. recalls GYN LAPAROSCOPY/LAPAROTOMY; Model/Item Number (Kit): 1. WRNM002-15Product may be potentially contaminated with a lubricant used during the manufacturing process.
-
Class IIofficial
AVID Medical, Inc. recalls ADULT HEART TRAY SHANDS JACKS; Model/Item Number (Kit): SHND419-08;Product may be potentially contaminated with a lubricant used during the manufacturing process.
-
Class IIofficial
AVID Medical, Inc. recalls ENDOVASCULAR PACK; Model/Item Number (Kit): VAPX014-04;Product may be potentially contaminated with a lubricant used during the manufacturing process.
-
Class IIofficial
AVID Medical, Inc. recalls CRANIOTOMY PACK; Model/Item Number (Kit): WRNM035-20;Product may be potentially contaminated with a lubricant used during the manufacturing process.
-
Class IIofficial
AVID Medical, Inc. recalls PACK OPEN HEART CFH; Model/Item Number (Kit): 1. UFCF004-06Product may be potentially contaminated with a lubricant used during the manufacturing process.
-
Class IIofficial
AVID Medical, Inc. recalls ROBOTIC PROSTATECTOMY PACK; Model/Item Number (Kit): NAMC423-11;Product may be potentially contaminated with a lubricant used during the manufacturing process.
-
Class IIofficial
AVID Medical, Inc. recalls LAPAROSCOPIC PROSTATECTOMY; Model/Item Number (Kit): NIHB0036-07;Product may be potentially contaminated with a lubricant used during the manufacturing process.
-
Class IIofficial
AVID Medical, Inc. recalls PLASTICS TRAY; Model/Item Number (Kit): EURO053;Product may be potentially contaminated with a lubricant used during the manufacturing process.
-
Class IIofficial
AVID Medical, Inc. recalls DAVINCI LAP/GYN; Model/Item Number (Kit): WRNM037-16;Product may be potentially contaminated with a lubricant used during the manufacturing process.
-
Class IIofficial
AVID Medical, Inc. recalls HEART PREP TRAY; Model/Item Number (Kit): MECV97W;Product may be potentially contaminated with a lubricant used during the manufacturing process.
-
Class IIofficial
AVID Medical, Inc. recalls UROLOGY MAJOR; Model/Item Number (Kit): WRNM058-10;Product may be potentially contaminated with a lubricant used during the manufacturing process.
-
Class IIofficial
AVID Medical, Inc. recalls MAJOR PACK; Model/Item Number (Kit): NIHB0018-11;Product may be potentially contaminated with a lubricant used during the manufacturing process.
-
Class IIofficial
AVID Medical, Inc. recalls GYN LAPAROSCOPY PACK; Model/Item Number (Kit): NAMC436-12;Product may be potentially contaminated with a lubricant used during the manufacturing process.
-
Class IIofficial
AVID Medical, Inc. recalls ABD HYSTERECTOMY; Model/Item Number (Kit): NAMC438-07;Product may be potentially contaminated with a lubricant used during the manufacturing process.
-
Class Iofficial
Bard Access Systems, Inc. recalls Power-Trialysis Short-Term Straight Dialysis Catheterthis is a downstream recall of class I recall RES 98777
-
Class Iofficial
Bard Access Systems, Inc. recalls DuoGlide Short-Term Straight Dialysis Catheterthis is a downstream recall of class I recall RES 98777