B. Braun Medical, Inc. recalls Infusomat UNIV 60 DROP PUMP SETPotential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime, resulting in patient injuries of adverse drug reaction and/or product malfunction with loss of medication/blood.
-
Class IIofficial
B. Braun Medical, Inc. recalls Outlook OUTLOOK PUMP SETPotential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime, resulting in patient injuries of adverse drug reaction and/or product malfunction with loss of medication/blood.
-
Class IIofficial
B. Braun Medical, Inc. recalls Infusomat UNIV. 15 DROP PUMP SETPotential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime, resulting in patient injuries of adverse drug reaction and/or product malfunction with loss of medication/blood.
-
Class IIofficial
B. Braun Medical, Inc. recalls OUTLOOK PUMP SET 3 CARESITE LADS 133 IN.-Used with an electrically-powered infusion pump for intravenous (IV)Potential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime, resulting in patient injuries of adverse drug reaction and/or product malfunction with loss of medication/blood.
-
Class IIofficial
ROi CPS LLC recalls Regard Kit containing the Microtek decanterDecanters may have pin holes and tears in sterile packaging that cannot be identified visually.
-
Class IIofficial
Abbott Diagnostics Technologies AS recalls Afinion 2 AnalyzerFive Afinion 2 (AF2) Analyzer were incorrectly market configured as Moderate Complexity and distributed to customers. Use of moderately complex assays by waived laboratories could result in erroneous results.
-
Class IIofficial
Medtronic Neuromodulation recalls Medtronic Handset with CommunicatorA subset of the handsets within the kit may not be able to complete the pairing process with the communicator upon initial setup.
-
Class IIofficial
Medical Depot Inc. recalls Nitro Glide Knee Walker-offers users recovering from a below-the-knee injury the ability to remain mobile andThe tiller separates from the base/front caster assembly. If the knee walker fails, a user may fall and sustain an injury
-
Class IIofficial
Medical Depot Inc. recalls Nitro Glide Knee Walker-offers users recovering from a below-the-knee injury the ability to remain mobile andThe tiller separates from the base/front caster assembly. If the knee walker fails, a user may fall and sustain an injury
-
Class IIofficial
B Braun Medical Inc recalls Brand Name: Perfusor Space Product Name: Perfusor Space Infusion Pump System Model/Catalog Number: 8713030U PrPerfusor Space Syringe Pump's instructions for use previously identified Monoject syringes as an option for use. However, changes to manufacturing to Cardinal Health Monoject syringes may result in issues with recognition, compatibility, and pump performance, such as overdose, underdose, delay in therapy, delay in occlusion alarms, and delay in feeding. Cardinal Health Monoject syringes are not compatible with the indicated pumps.
-
Class IIofficial
B Braun Medical Inc recalls Brand Name: Perfusor Space Product Name: Perfusor Space Syringe PumpPerfusor Space Syringe Pump's instructions for use previously identified Monoject syringes as an option for use. However, changes to manufacturing to Cardinal Health Monoject syringes may result in issues with recognition, compatibility, and pump performance, such as overdose, underdose, delay in therapy, delay in occlusion alarms, and delay in feeding. Cardinal Health Monoject syringes are not compatible with the indicated pumps.
-
Class IIofficial
B Braun Medical Inc recalls Brand Name: Perfusor Space Product Name: Perfusor PCA Syringe Pump Model/Catalog Number: 8713080U Product DescPerfusor Space Syringe Pump's instructions for use previously identified Monoject syringes as an option for use. However, changes to manufacturing to Cardinal Health Monoject syringes may result in issues with recognition, compatibility, and pump performance, such as overdose, underdose, delay in therapy, delay in occlusion alarms, and delay in feeding. Cardinal Health Monoject syringes are not compatible with the indicated pumps.
-
Class IIofficial
B Braun Medical Inc recalls Brand Name: Perfusor Space Product Name: PERFUSOR SPACE ACM-KIT Model/Catalog Number: 8713040U Product DescripPerfusor Space Syringe Pump's instructions for use previously identified Monoject syringes as an option for use. However, changes to manufacturing to Cardinal Health Monoject syringes may result in issues with recognition, compatibility, and pump performance, such as overdose, underdose, delay in therapy, delay in occlusion alarms, and delay in feeding. Cardinal Health Monoject syringes are not compatible with the indicated pumps.
-
Class IIofficial
B Braun Medical Inc recalls Brand Name: Perfusor Space Product Name: Perfusor Space Syringe PumpPerfusor Space Syringe Pump's instructions for use previously identified Monoject syringes as an option for use. However, changes to manufacturing to Cardinal Health Monoject syringes may result in issues with recognition, compatibility, and pump performance, such as overdose, underdose, delay in therapy, delay in occlusion alarms, and delay in feeding. Cardinal Health Monoject syringes are not compatible with the indicated pumps.
-
Class IIofficial
GE Medical Systems, LLC recalls SIGNA ArchitectGE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
-
Class IIofficial
GE Medical Systems, LLC recalls SIGNA PioneerGE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
-
Class IIofficial
GE Medical Systems, LLC recalls SIGNA HeroGE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
-
Class IIofficial
GE Medical Systems, LLC recalls SIGNA PremierGE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
-
Class IIofficial
GE Medical Systems, LLC recalls 3.0T SIGNA HDxt FamilyGE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
-
Class IIofficial
GE Medical Systems, LLC recalls Discovery MR750 3.0TGE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
-
Class IIofficial
GE Medical Systems, LLC recalls Discovery MR750w 3.0TGE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
-
Class IIofficial
GE Medical Systems, LLC recalls SIGNA PET/MRGE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
-
Class IIofficial
GE Medical Systems, LLC recalls Discovery MR450 1.5TGE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
-
Class IIofficial
GE Medical Systems, LLC recalls Optima MR450w 1.5TGE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
-
Class IIofficial
GE Medical Systems, LLC recalls SIGNA ArtistGE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
-
Class IIofficial
GE Medical Systems, LLC recalls SIGNA VoyagerGE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
-
Class IIofficial
GE Medical Systems, LLC recalls 1.5T SIGNA HDxt FamilyGE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
-
Class IIofficial
GE Medical Systems, LLC recalls SIGNA CreatorGE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
-
Class IIofficial
GE Medical Systems, LLC recalls SIGNA ExplorerGE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
-
Class IIofficial
GE Medical Systems, LLC recalls SIGNA MR380GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
-
Class IIofficial
GE Medical Systems, LLC recalls SIGNA MR355GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
-
Class IIofficial
GE Medical Systems, LLC recalls SIGNA MR360GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
-
Class IIofficial
GE Medical Systems, LLC recalls Brivo MR355GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
-
Class IIofficial
GE Medical Systems, LLC recalls Optima MR360GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
-
Class IIofficial
GE Medical Systems, LLC recalls SIGNA PrimeGE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
-
Class IIofficial
GE Medical Systems, LLC recalls SIGNA VictorGE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
-
Class IIofficial
GE Medical Systems, LLC recalls SIGNA UHPGE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
-
Class IIofficial
GE Medical Systems, LLC recalls Discovery MR950GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
-
Class IIofficial
GE Medical Systems, LLC recalls SIGNA 7.0TGE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
-
Class IIofficial
GE Medical Systems, LLC recalls SIGNA MAGNUSGE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
-
Class IIofficial
Baxter Healthcare Corporation recalls Baxter SIGMA Spectrum Infusion PumpImproperly performed testing prior to release
-
Class IIofficial
Beckman Coulter, Inc. recalls Discrete photometric chemistry analyzer for clinical use.Beckman Coulter, Inc. has identified a rise in reagent carousel motion errors in Access 2 Immunoassay System Analyzers that appear to be caused by misaligned index wheels housed within the reagent carousel module. Customers affected by this issue may observe an increase in reagent carousel motion errors or QSMC errors that occur while unloading or loading reagent packs during system initialization and sample processing, which causes a non-operational system status and cancels tests; and patient results may be delayed if these errors cancel test processing.
-
Class IIofficial
Argon Medical Devices, Inc recalls L-CATH PICC S/L Peripherally Inserted Central CatheterPeripherally Inserted Central Catheter (PICC) is labeled 26G/1.9Fr x 30cm but package contains incorrect 26G/1.9Fr x 50cm PICC, which could lead to procedural delay.
-
Class IIofficial
Abbott Laboratories recalls Alinity hq Analyzer1)If Complete Blood Count with differential and reticulocyte(CBC+Diff+Retic) sample is run, then immediately following on same rack CBC+Diff sample is run, then falsely low red blood cell count may occur, generating falsely high Mean Cell Hemoglobin(MCH)/MCHC and falsely low hematocrit results. 2)Cell events may be incorrectly counted as basophil(BASO), resulting in increased BASO and %BASO counts
-
Class Iofficial
Totally Cool, Inc. recalls Friendly's Celebration Ice Cream Cake 60 fl ozPossible contamination with Listeria monocytogenes
-
Class IIofficial
Totally Cool, Inc. recalls Friendly's Strawberry Krunch Ice Cream Cake 40 fl ozPossible contamination with Listeria monocytogenes
-
Class IIofficial
Totally Cool, Inc. recalls Abilyn s Frozen Bakery Ice Cream Cake Vanilla & Chocolate 6"Possible contamination with Listeria monocytogenes
-
Class IIofficial
Totally Cool, Inc. recalls Abilyn s Frozen Bakery Ice Cream Cake Cookies 'n Cream 6"Possible contamination with Listeria monocytogenes
-
Class IIofficial
Totally Cool, Inc. recalls Abilyn s Frozen Bakery Ice Cream Cake For the Love of Chocolate 6"Possible contamination with Listeria monocytogenes
-
Class IIofficial
Totally Cool, Inc. recalls Hershey s Ice Cream Vanilla & Chocolate Flavored Ice Cream Cake 38 fl oz 7"Possible contamination with Listeria monocytogenes