Class IIofficial
Philips Medical Systems Nederland B.v. recalls Philips Azurion 7M20During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.
-
Class IIofficial
Abiomed, Inc. recalls Impella ConnectThe following features have been disabled from the web-based portal because the FDA has not evaluated these features for safety and effectiveness: Notifications via email feature; AIC alarm color on case tile feature; Sort case tiles by AIC alarm color feature; Pump metric display on case tile feature
-
Class IIofficial
Baxter Healthcare Corporation recalls Baxter Flo-ResterCardiovascular Surgery marketing brochure (US-AS46-220001) that was distributed for the products listed below contains inaccurate content that does not align with some of the products Instructions for Use (IFU) or Prescribing Information (PI).
-
Class IIofficial
Baxter Healthcare Corporation recalls Baxter Coseal PremixCardiovascular Surgery marketing brochure (US-AS46-220001) that was distributed for the products listed below contains inaccurate content that does not align with some of the products Instructions for Use (IFU) or Prescribing Information (PI).
-
Class IIofficial
Baxter Healthcare Corporation recalls Baxter Ostene Hemostasis MaterialCardiovascular Surgery marketing brochure (US-AS46-220001) that was distributed for the products listed below contains inaccurate content that does not align with some of the products Instructions for Use (IFU) or Prescribing Information (PI).
-
Class IIofficial
Baxter Healthcare Corporation recalls Baxter Vascular ProbeCardiovascular Surgery marketing brochure (US-AS46-220001) that was distributed for the products listed below contains inaccurate content that does not align with some of the products Instructions for Use (IFU) or Prescribing Information (PI).
-
Class IIofficial
Baxter Healthcare Corporation recalls Baxter Preveleak Surgical SealantCardiovascular Surgery marketing brochure (US-AS46-220001) that was distributed for the products listed below contains inaccurate content that does not align with some of the products Instructions for Use (IFU) or Prescribing Information (PI).
-
Class IIofficial
Baxter Healthcare Corporation recalls Baxter Floseal Hemostatic Matrix Fast PrepCardiovascular Surgery marketing brochure (US-AS46-220001) that was distributed for the products listed below contains inaccurate content that does not align with some of the products Instructions for Use (IFU) or Prescribing Information (PI).
-
Class IIofficial
Baxter Healthcare Corporation recalls Baxter FloSeal Hemostatic Matrix RECOTHROMCardiovascular Surgery marketing brochure (US-AS46-220001) that was distributed for the products listed below contains inaccurate content that does not align with some of the products Instructions for Use (IFU) or Prescribing Information (PI).
-
Class IIofficial
Baxter Healthcare Corporation recalls Baxter Flo-Thru Intraluminal ShuntCardiovascular Surgery marketing brochure (US-AS46-220001) that was distributed for the products listed below contains inaccurate content that does not align with some of the products Instructions for Use (IFU) or Prescribing Information (PI).
-
Class IIofficial
Baxter Healthcare Corporation recalls Peri-Guard Repair PatchCardiovascular Surgery marketing brochure (US-AS46-220001) that was distributed for the products listed below contains inaccurate content that does not align with some of the products Instructions for Use (IFU) or Prescribing Information (PI).
-
Class IIofficial
Baxter Healthcare Corporation recalls Peri-Guard Repair PatchCardiovascular Surgery marketing brochure (US-AS46-220001) that was distributed for the products listed below contains inaccurate content that does not align with some of the products Instructions for Use (IFU) or Prescribing Information (PI).
-
Class IIofficial
Baxter Healthcare Corporation recalls Vascu-Guard Peripheral Vascular Patch (US)Cardiovascular Surgery marketing brochure (US-AS46-220001) that was distributed for the products listed below contains inaccurate content that does not align with some of the products Instructions for Use (IFU) or Prescribing Information (PI).
-
Class IIofficial
Baxter Healthcare Corporation recalls Baxter FloSeal Hemostatic Matrix NLCardiovascular Surgery marketing brochure (US-AS46-220001) that was distributed for the products listed below contains inaccurate content that does not align with some of the products Instructions for Use (IFU) or Prescribing Information (PI).
-
Class IIofficial
Baxter Healthcare Corporation recalls PERI-GUARD SUPPLECardiovascular Surgery marketing brochure (US-AS46-220001) that was distributed for the products listed below contains inaccurate content that does not align with some of the products Instructions for Use (IFU) or Prescribing Information (PI).
-
Class IIofficial
Baxter Healthcare Corporation recalls Vascu-Guard Peripheral Vascular Patch (GLOBAL)Cardiovascular Surgery marketing brochure (US-AS46-220001) that was distributed for the products listed below contains inaccurate content that does not align with some of the products Instructions for Use (IFU) or Prescribing Information (PI).
-
Class IIofficial
Accure Acne, Inc recalls Accure Laser SystemLaser system's spot tracking system may become misaligned resulting in the possible risk of delivery of higher-than-expected laser energy to epidermal tissue, which may cause blistering.
-
Class IIofficial
Stryker, Inc. recalls Stryker NeurovascularThe microcatheter IFU contains an intended use that has not been evaluated by the FDA.
-
Class IIofficial
Stryker, Inc. recalls Stryker NeurovascularThe microcatheter IFU contains an intended use that has not been evaluated by the FDA.
-
Class IIofficial
Ambu Inc. recalls Ambu aView 2 A MonitorInstructions for Use have been updated to add the following warning: Use only M4 screws with the length of 14 16 mm when mounting Ambu aView 2 Advance on a VESA interface. Using longer screw lengths will penetrate the lithium-ion battery and result in a fire hazard and battery leakage which can cause severe burns, smoke inhalation and skin irritation. Using shorter screw lengths could result in unsecure device fastening.
-
Class IIofficial
Oakworks Inc recalls Ultrasound EA (Ergonomic Advantage) table versions; G2 (GENERAL 2) -An Ultrasound examination tableBackrest platform stress fracture - the table is not operational in an inclined position
-
Class IIofficial
Oakworks Inc recalls Ultrasound EA (Ergonomic Advantage) table versions; G3 (General 3 Section) Echocardiography -An Ultrasound exaBackrest platform stress fracture - the table is not operational in an inclined position
-
Class IIofficial
Oakworks Inc recalls Ultrasound EA (Ergonomic Advantage) table versions; G3 Multi-Specialty Echocardiography and Multi-Specialty wiBackrest platform stress fracture - the table is not operational in an inclined position
-
Class IIofficial
Oakworks Inc recalls Ultrasound EA (Ergonomic Advantage) table version; V2 (Vascular 2 Section) Echocardiography -An Ultrasound exaBackrest platform stress fracture - the table is not operational in an inclined position
-
Class IIofficial
Shirakawa Olympus Co., Ltd. recalls Model No. CYF-5Olympus has updated the instructions for use: Reusable cleaning brushes have been removed from the manual cleaning. Single use brushes remain compatible.
-
Class IIofficial
Shirakawa Olympus Co., Ltd. recalls Model No. CYF-5ROlympus has updated the instructions for use: Reusable cleaning brushes have been removed from the manual cleaning. Single use brushes remain compatible.
-
Class IIofficial
Shirakawa Olympus Co., Ltd. recalls Model No. BF-P190Olympus has updated the instructions for use: Reusable cleaning brushes have been removed from the manual cleaning. Single use brushes remain compatible.
-
Class IIofficial
Shirakawa Olympus Co., Ltd. recalls Model No. BF-XP190Olympus has updated the instructions for use: Reusable cleaning brushes have been removed from the manual cleaning. Single use brushes remain compatible.
-
Class IIofficial
Shirakawa Olympus Co., Ltd. recalls Model No. ENF-XPOlympus has updated the instructions for use: Reusable cleaning brushes have been removed from the manual cleaning. Single use brushes remain compatible.
-
Class IIofficial
Shirakawa Olympus Co., Ltd. recalls Model No. CYF-VHOlympus has updated the instructions for use: Reusable cleaning brushes have been removed from the manual cleaning. Single use brushes remain compatible.
-
Class IIofficial
Shirakawa Olympus Co., Ltd. recalls Model No. CYF-VHROlympus has updated the instructions for use: Reusable cleaning brushes have been removed from the manual cleaning. Single use brushes remain compatible.
-
Class IIofficial
Shirakawa Olympus Co., Ltd. recalls Model No. CYF-V2Olympus has updated the instructions for use: Reusable cleaning brushes have been removed from the manual cleaning. Single use brushes remain compatible.
-
Class IIofficial
Shirakawa Olympus Co., Ltd. recalls Model No. CYF-V2ROlympus has updated the instructions for use: Reusable cleaning brushes have been removed from the manual cleaning. Single use brushes remain compatible.
-
Class IIofficial
Olympus Corporation of the Americas recalls Soltive SuperPulsed Laser System TFL Fiber StripperThe strippers and cleavers are provided non-sterile and labelled as reusable/autoclavable. Olympus does not have validated cleaning and sterilization instructions. Use of a non-sterile fiber stripper or cleaver on a sterile fiber poses a risk of contamination.
-
Class IIofficial
Olympus Corporation of the Americas recalls Soltive SuperPulsed Laser System TFL FIBER CLEAVERThe strippers and cleavers are provided non-sterile and labelled as reusable/autoclavable. Olympus does not have validated cleaning and sterilization instructions. Use of a non-sterile fiber stripper or cleaver on a sterile fiber poses a risk of contamination.
-
Class IIofficial
Olympus Corporation of the Americas recalls SOLTIVE SuperPulsed Laser FibersAffected products may potentially have a breached sterile pouch seal. A breached sterile pouch seal may lead to minor procedure delays (if noticed prior to use) or patient infections (if not noticed prior to use in patient).
-
Class Iofficial
Cardinal Health 200, LLC recalls Argyle" Suction Catheter Tray with Chimney Valve Sterile Saline 18 Fr/Ch (6.0mm); Argyle" Suction Catheter TraConvenience kits were manufactured and distributed with components which were subsequently recalled by Nurse Assist.
-
Class Iofficial
Cardinal Health 200, LLC recalls TRACH KIT W/HYDROGEN & SALIConvenience kits were manufactured and distributed with components which were subsequently recalled by Nurse Assist.
-
Class Iofficial
Cardinal Health 200, LLC recalls Dover" 100% SILICONE 400 ML U/M FOLEY TRAYConvenience kits were manufactured and distributed with components which were subsequently recalled by Nurse Assist.
-
Class Iofficial
Cardinal Health 200, LLC recalls Dover URINE METER ADD-A-FOLEY TRAYConvenience kits were manufactured and distributed with components which were subsequently recalled by Nurse Assist.
-
Class Iofficial
Cardinal Health 200, LLC recalls Dover" HYDROGEL COATED LATEX 400 ML U/M FOLEY TRAYConvenience kits were manufactured and distributed with components which were subsequently recalled by Nurse Assist.
-
Class Iofficial
Cardinal Health 200, LLC recalls Dover" 100% SILICONE 400 ML U/M FOLEY TRAYConvenience kits were manufactured and distributed with components which were subsequently recalled by Nurse Assist.
-
Class Iofficial
Cardinal Health 200, LLC recalls Dover" IRRIGATION 1200 CC TRAYConvenience kits were manufactured and distributed with components which were subsequently recalled by Nurse Assist.
-
Class Iofficial
Cardinal Health 200, LLC recalls Dover" 100% SILICONE FOLEY TRAYConvenience kits were manufactured and distributed with components which were subsequently recalled by Nurse Assist.
-
Class Iofficial
Cardinal Health 200, LLC recalls Dover" 100% SILICONE 400ML U/M TEMP SENSE FOLEY TRAYConvenience kits were manufactured and distributed with components which were subsequently recalled by Nurse Assist.
-
Class Iofficial
Cardinal Health 200, LLC recalls Dover" 100% SILICONE FOLEY TRAYConvenience kits were manufactured and distributed with components which were subsequently recalled by Nurse Assist.
-
Class IIofficial
Maquet Cardiovascular, LLC recalls Vasoview Hemopro Endoscopic Vessel Harvesting SystemRisk of polyimide particulate matter being present on the endoscopic harvesting vessel system. This issue may result in polyimide particulate matter being introduced into the arterial conduit for patients undergoing a coronary artery bypass graft (CABG) or peripheral bypass graft surgery.
-
Class IIofficial
Maquet Cardiovascular, LLC recalls Vasoview Hemopro Endoscopic Vessel Harvesting SystemRisk of polyimide particulate matter being present on the endoscopic harvesting vessel system. This issue may result in polyimide particulate matter being introduced into the arterial conduit for patients undergoing a coronary artery bypass graft (CABG) or peripheral bypass graft surgery.
-
Class IIofficial
Olympus Corporation of the Americas recalls SOLTIVE SuperPulsed Laser Fibers Model Numbers TFL-FBX150SOlympus identified inconsistencies in the Instructions for Use regarding cleaning the fiber tip and fiber cleaving and stripping instructions. Following the current instructions presents the potential for the fiber to break during use, which may cause burns, procedure delays, prolonged surgery, and foreign body in patient.
-
Class IIofficial
Howmedica Osteonics Corp. recalls SCORPIO RECESSED PATELLA- Intended for Knee Replacement Part Number: 3044-0026Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.