Becton Dickinson & Company recalls BD Extension Sets (Product NameAffected infusion sets labeled as free of DEHP actually do contain di(2-ethylhexyl) phthalate (DEHP).
-
Class IIofficial
SunMed Holdings, LLC recalls VentlabA backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.
-
Class IIofficial
SunMed Holdings, LLC recalls VentlabA backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.
-
Class IIofficial
SunMed Holdings, LLC recalls VentlabA backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.
-
Class IIofficial
SunMed Holdings, LLC recalls VentlabA backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.
-
Class IIofficial
SunMed Holdings, LLC recalls VentlabA backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.
-
Class IIofficial
SunMed Holdings, LLC recalls VentlabA backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.
-
Class IIofficial
SunMed Holdings, LLC recalls VentlabA backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.
-
Class IIofficial
SunMed Holdings, LLC recalls Curaplex manual resuscitators with integrated manometerA backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.
-
Class IIofficial
SunMed Holdings, LLC recalls Medline manual resuscitators with integrated manometerA backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.
-
Class IIofficial
Randox Laboratories Ltd. recalls RANDOX Total Bilirubin.Randox has had reports of elevated patient results using Plasma (lithium heparin) samples with Total Bilirubin reagent (TBil) Vanadate Oxidation Method.
-
Class IIofficial
Mc3 Inc recalls NautilusTM ECMO Oxygenator with Balance Biosurface-used in extracorporeal life support procedures to oxygenateMay have increased water path restriction within the heat exchanger, leading to degraded heat exchange performance during use. The reduction of water flow and reduced heat exchange performance may be experienced immediately upon priming or may worsen over the duration of use. The potential harm as stated in the IFU includes hypothermia due to inadequate heat transfer
-
Class IIofficial
Mc3 Inc recalls NautilusTM Smart ECMO Module with Balance Biosurface-used in extracorporeal life support procedures to oxygenaMay have increased water path restriction within the heat exchanger, leading to degraded heat exchange performance during use. The reduction of water flow and reduced heat exchange performance may be experienced immediately upon priming or may worsen over the duration of use. The potential harm as stated in the IFU includes hypothermia due to inadequate heat transfer
-
Class IIofficial
The Cookie Factory LLC recalls Stewarts Chocolate Chip Cookies; 6 oz package; packaged in a flexible plastic clamshell type packageProduct contains undeclared macadamia nuts.
-
Class Iofficial
Maquet Medical Systems USA recalls CARDIOHELP Emergency DriveProduct removal due to possible blocking or impairment of the CARDIOHELP Emergency Drive. This issue can lead to the patient being exposed to inappropriate low blood flow. Potential associated harms are ischemia and hypoxia.
-
Class IIofficial
Abiomed, Inc. recalls Automated Impella Controller (AIC)Pump not detected as connected to controller due to software issue.
-
Class IIofficial
Abiomed, Inc. recalls Automated Impella Controller (AIC)Pump not detected as connected to controller due to software issue.
-
Class IIofficial
Fresenius Medical Care Holdings, Inc. recalls Stay Safe CapUpdating the labeling to include additional warnings and contraindications for potential side effects.
-
Class IIofficial
Ortho-Clinical Diagnostics, Inc. recalls VITROS 3600 Immunodiagnostic System Product Code 6802783 and Refurbished Product Code 6802914 Running SoftwareDuring planned monitoring of data post-release of Software Version 3.8.0 an increased occurrence of software issues relating to unresponsive subsystems, unresponsive user interface (sometimes referred to as a screen freeze), and/or printer disconnections was identified. The consequence of these issues would be a delay in reporting results due to the required system shutdowns to recover the system. This delay in reporting results could occur during any assay, including which the speed of the result is deemed critical or STAT and could result in a delay of treatment in a patient. Due to the release strategy of this software version, the user could not update the system but a field engineer had to manually install each affected system.
-
Class IIofficial
Ortho-Clinical Diagnostics, Inc. recalls VITROS 4600 Chemistry System Product Code 6802445 and Refurbished Product Code 6900440 Running Software VersioDuring planned monitoring of data post-release of Software Version 3.8.0 an increased occurrence of software issues relating to unresponsive subsystems, unresponsive user interface (sometimes referred to as a screen freeze), and/or printer disconnections was identified. The consequence of these issues would be a delay in reporting results due to the required system shutdowns to recover the system. This delay in reporting results could occur during any assay, including which the speed of the result is deemed critical or STAT and could result in a delay of treatment in a patient. Due to the release strategy of this software version, the user could not update the system but a field engineer had to manually install each affected system.
-
Class IIofficial
Ortho-Clinical Diagnostics, Inc. recalls VITROS 5600 Integrated System Product Code 6802413 and Refurbished Product Code 6802915 Running Software VersiDuring planned monitoring of data post-release of Software Version 3.8.0 an increased occurrence of software issues relating to unresponsive subsystems, unresponsive user interface (sometimes referred to as a screen freeze), and/or printer disconnections was identified. The consequence of these issues would be a delay in reporting results due to the required system shutdowns to recover the system. This delay in reporting results could occur during any assay, including which the speed of the result is deemed critical or STAT and could result in a delay of treatment in a patient. Due to the release strategy of this software version, the user could not update the system but a field engineer had to manually install each affected system.
-
Class IIofficial
Ortho-Clinical Diagnostics, Inc. recalls VITROS XT 3400 Chemistry System Product Code 6844458 Running Software Version 3.8.0During planned monitoring of data post-release of Software Version 3.8.0 an increased occurrence of software issues relating to unresponsive subsystems, unresponsive user interface (sometimes referred to as a screen freeze), and/or printer disconnections was identified. The consequence of these issues would be a delay in reporting results due to the required system shutdowns to recover the system. This delay in reporting results could occur during any assay, including which the speed of the result is deemed critical or STAT and could result in a delay of treatment in a patient. Due to the release strategy of this software version, the user could not update the system but a field engineer had to manually install each affected system.
-
Class IIofficial
Ortho-Clinical Diagnostics, Inc. recalls VITROS XT 7600 Integrated System Product Code 6844461 Running Software Version 3.8.0During planned monitoring of data post-release of Software Version 3.8.0 an increased occurrence of software issues relating to unresponsive subsystems, unresponsive user interface (sometimes referred to as a screen freeze), and/or printer disconnections was identified. The consequence of these issues would be a delay in reporting results due to the required system shutdowns to recover the system. This delay in reporting results could occur during any assay, including which the speed of the result is deemed critical or STAT and could result in a delay of treatment in a patient. Due to the release strategy of this software version, the user could not update the system but a field engineer had to manually install each affected system.
-
Class IIofficial
Beaver Visitec International, Inc. recalls Xstar Kojo Safety Slit KnifeOne lot of Part Number 373025, Xstar¿ Kojo Safety Slit Knife 2.5mm 45¿ bevel up, incorrectly contains a bevel down blade instead of a bevel up blade.
-
Class Iofficial
Greenhead Lobster Products LLC recalls Frozen and fresh RTE Lobster meat (1 lb. packages to 12 lb. case) 1. Greenhead Lobster HPP Fresh Cooked Claw aLobster meat may be contaminated with Listeria monocytogenes.
-
Class IIofficial
Philips Ultrasound, Inc. recalls Philips EPIQ Diagnostic Ultrasound System Models: EPIQ 7G(model # 795200)Their is the potential that diagnostic ultrasound systems operating with software version 10.0 may present a delayed image instead of a real time image.
-
Class IIofficial
Philips Ultrasound, Inc. recalls Philips Affiniti 70Their is the potential that diagnostic ultrasound systems operating with software version 10.0 may present a delayed image instead of a real time image.
-
Class IIofficial
Philips North America Llc recalls Big Bore RT with software version V4.8.0.10421- Model: 728242 CT scanner and simulator designed for radiationFor oncology users: If the user performed off-set reconstruction on CT device a shift could be observed on the contouring area of the primary and secondary images sets when using image fusion (on TPS) and incorrect coordinates may be sent to TPS system when using absolute patient marking. This may lead to incorrect radiation therapy planning and possible growth or spread of cancer due to incorrect early treatment regimen. For radiology users: If the user performed off-set reconstruction on CT devices, after multiple acquisitions, the Relate Position could be inaccurate. This could lead to incorrect diagnosis and treatment.
-
Class IIofficial
Philips North America Llc recalls Brilliance CT Big Bore with software version V4.8.0.10421 Model:728244 CT scanner and simulator designed for rFor oncology users: If the user performed off-set reconstruction on CT device a shift could be observed on the contouring area of the primary and secondary images sets when using image fusion (on TPS) and incorrect coordinates may be sent to TPS system when using absolute patient marking. This may lead to incorrect radiation therapy planning and possible growth or spread of cancer due to incorrect early treatment regimen. For radiology users: If the user performed off-set reconstruction on CT devices, after multiple acquisitions, the Relate Position could be inaccurate. This could lead to incorrect diagnosis and treatment.
-
Class IIofficial
Philips North America Llc recalls Brilliance CT Big Bore Oncology with software version V4.8.0.10421 Model:728243 CT scanner and simulator desigFor oncology users: If the user performed off-set reconstruction on CT device a shift could be observed on the contouring area of the primary and secondary images sets when using image fusion (on TPS) and incorrect coordinates may be sent to TPS system when using absolute patient marking. This may lead to incorrect radiation therapy planning and possible growth or spread of cancer due to incorrect early treatment regimen. For radiology users: If the user performed off-set reconstruction on CT devices, after multiple acquisitions, the Relate Position could be inaccurate. This could lead to incorrect diagnosis and treatment.
-
Class IIofficial
Philips North America recalls Spectral CT on RailsWhen preparing for CCT Fluoroscopy mode and activating the CCT foot pedal simultaneously, the operator may experience a software crash. This would result in the system needing to be restarted.
-
Class IIofficial
Philips North America recalls Spectral CT on RailsThis recall is part of a 2-issue recall for the same system. When using the IVC needle position button, on the IVC control box during a series of scans, the plan box may not move to the desired needle position. This issue could result in the gantry moving to a location the operator does not expect. If this occurs, it presents a risk that the scan could occur at an incorrect location, or that the unintended motion of the gantry could cause a collision with the user or needle. The other part of the recall involves a software crash when the CCT foot pedal is activated (as described in RES ID93363).
-
Class IIofficial
Biomet, Inc. recalls Compress Device Segmental Anchor PlugAnchor plugs potentially have metal burrs in the transverse holes of the device, preventing the drill or pins from passing through, may result in extension of surgery
-
Class IIofficial
Biomet, Inc. recalls Compress Device Segmental Anchor PlugAnchor plugs potentially have metal burrs in the transverse holes of the device, preventing the drill or pins from passing through, may result in extension of surgery
-
Class IIofficial
Biomet, Inc. recalls Compress Device Segmental Anchor PlugAnchor plugs potentially have metal burrs in the transverse holes of the device, preventing the drill or pins from passing through, may result in extension of surgery
-
Class IIofficial
Biomet, Inc. recalls Compress Device Segmental Anchor PlugAnchor plugs potentially have metal burrs in the transverse holes of the device, preventing the drill or pins from passing through, may result in extension of surgery
-
Class IIofficial
Biomet, Inc. recalls Compress Device Segmental Anchor PlugAnchor plugs potentially have metal burrs in the transverse holes of the device, preventing the drill or pins from passing through, may result in extension of surgery
-
Class IIofficial
Biomet, Inc. recalls Compress Device Segmental Anchor PlugAnchor plugs potentially have metal burrs in the transverse holes of the device, preventing the drill or pins from passing through, may result in extension of surgery
-
Class IIofficial
Biomet, Inc. recalls Compress Device Segmental Anchor PlugAnchor plugs potentially have metal burrs in the transverse holes of the device, preventing the drill or pins from passing through, may result in extension of surgery
-
Class IIofficial
Biomet, Inc. recalls Compress Device Segmental Anchor PlugAnchor plugs potentially have metal burrs in the transverse holes of the device, preventing the drill or pins from passing through, may result in extension of surgery
-
Class IIofficial
Biomet, Inc. recalls Compress Device Short Anchor PlugAnchor plugs potentially have metal burrs in the transverse holes of the device, preventing the drill or pins from passing through, may result in extension of surgery
-
Class IIofficial
Biomet, Inc. recalls Compress Device Short Anchor PlugAnchor plugs potentially have metal burrs in the transverse holes of the device, preventing the drill or pins from passing through, may result in extension of surgery
-
Class IIofficial
Biomet, Inc. recalls Compress Device Short Anchor PlugAnchor plugs potentially have metal burrs in the transverse holes of the device, preventing the drill or pins from passing through, may result in extension of surgery
-
Class IIofficial
Biomet, Inc. recalls Compress Device Short Anchor PlugAnchor plugs potentially have metal burrs in the transverse holes of the device, preventing the drill or pins from passing through, may result in extension of surgery
-
Class IIofficial
Biomet, Inc. recalls Compress Device Short Anchor PlugAnchor plugs potentially have metal burrs in the transverse holes of the device, preventing the drill or pins from passing through, may result in extension of surgery
-
Class IIofficial
Biomet, Inc. recalls Compress Device Short Anchor PlugAnchor plugs potentially have metal burrs in the transverse holes of the device, preventing the drill or pins from passing through, may result in extension of surgery
-
Class IIofficial
Biomet, Inc. recalls Compress Device Short Anchor PlugAnchor plugs potentially have metal burrs in the transverse holes of the device, preventing the drill or pins from passing through, may result in extension of surgery
-
Class IIofficial
Biomet, Inc. recalls Compress Device Short Anchor PlugAnchor plugs potentially have metal burrs in the transverse holes of the device, preventing the drill or pins from passing through, may result in extension of surgery
-
Class IIofficial
Biomet, Inc. recalls Compress Device Short Anchor PlugAnchor plugs potentially have metal burrs in the transverse holes of the device, preventing the drill or pins from passing through, may result in extension of surgery
-
Class IIofficial
Biomet, Inc. recalls Compress Device Short Anchor PlugAnchor plugs potentially have metal burrs in the transverse holes of the device, preventing the drill or pins from passing through, may result in extension of surgery