Class IIofficial
Terex Utilities recalls Power Over Fiber Base ModuleDevice not certified - no laser product report was filed with CDRH prior to distribution. Product label incorrectly states device is "Class IV", although Class 1 is appropriate for device. Improperly labeled - omitted labels required under 21 CFR 1040.10 / IEC 60825-1 via LN#56. Missing certification and proper manufacturer's ID label. Was not accompanied by compliant User Information. Formalized quality control procedures and record keeping of QC testing was not performed to IEC 60825-1 nor 21 CFR 1040.10 standards. Sales information did not convey required class designation and warning statement
-
Class Iofficial
Datex-Ohmeda Inc recalls Giraffe Stand-alone Infant Resuscitation System; Powered emergency ventilator (resuscitator) Product identifieBlender knob shaft can loosen, which can affect the delivered oxygen concentration. The system can deliver an oxygen concentration that does not match the intended concentration set by the user, potentially resulting in hypoxia or hyperoxia in the patient.
-
Class Iofficial
Datex-Ohmeda Inc recalls Giraffe Warmer with Integrated Resuscitation System (iRes)Blender knob shaft can loosen, which can affect the delivered oxygen concentration. The system can deliver an oxygen concentration that does not match the intended concentration set by the user, potentially resulting in hypoxia or hyperoxia in the patient.
-
Class Iofficial
Datex-Ohmeda Inc recalls Panda Freestanding WarmerBlender knob shaft can loosen, which can affect the delivered oxygen concentration. The system can deliver an oxygen concentration that does not match the intended concentration set by the user, potentially resulting in hypoxia or hyperoxia in the patient.
-
Class IIofficial
AMD Medicom Inc. recalls Ritmed Neurological SpongesNeurological sponges used to moisten or absorb biological fluid during neurological surgery and may also be used to protect the brain during incision, may not consistently meet endotoxin specification for devices that contact the cerebrospinal fluid. Elevated endotoxin levels may lead to neurovascular adverse events, febrile response, and/or local transient inflammation, hypotension, or nausea.
-
Class IIofficial
AMD Medicom Inc. recalls Ritmed Neurological SpongesNeurological sponges used to moisten or absorb biological fluid during neurological surgery and may also be used to protect the brain during incision, may not consistently meet endotoxin specification for devices that contact the cerebrospinal fluid. Elevated endotoxin levels may lead to neurovascular adverse events, febrile response, and/or local transient inflammation, hypotension, or nausea.
-
Class IIofficial
Merit Medical Systems, Inc. recalls Splashwire Hydrophilic Guide Wire: REF: MSWSTDA35150 UDI-DI codes: 00884450529645 00884450529669 0088445052969Due to unsealed portions of pouches.
-
Class IIofficial
Ventec Life Systems, Inc. recalls VOCSN+Pro packageRespiratory system intended to provide continuous/intermittent ventilator support, may have a damaged back-up battery supporting backup alarm function, which may result in the backup alarm not activating during loss of primary power and absence/depletion of backup power sources, which may result in delayed caregiver awareness of a hazardous condition.
-
Class IIofficial
GE Medical Systems SCS recalls GE Healthcare AW Server 3.2 ext.6.5; Radiological Image Processing SystemGE HealthCare has become aware of a context synchronization issue in AW Server 3.2 ext. 6.5. When a user selects a patient or exam in the AW Server Web Client worklist and launches an interactive application (e.g. Volume Viewer), the application may open the previous patient's exam instead of the intended one. When this issue occurs, there is no system warning or error notification. If the error is not recognized, a clinical user could review, interpret, or report images for the wrong patient which could lead to misdiagnosis, incorrect clinical decisions, resulting in delayed or incorrect treatment.
-
Class IIofficial
Think Surgical, Inc. recalls TMINI Femur Tracking Array REF 109003 UDI-DI code: M9461090030 TMINI Tibia Tracking Array REF 109004 UDI-DI coDue to manufacturing/design issue where the knob on the tracking array may loosen causing potential misalignment of the array.
-
Class IIofficial
Instrumentation Laboratory recalls GEM Premier 5000 with IQM2 Model/Catalog Number: 00024019255 and GEM Premier 7000 with iQM3 Model/Catalog NumbPotential for Bar Code Reader (BCR) wands, used with the GEM Premier 5000 and GEM Premier 7000 systems, to assign patient results to an incorrect patient ID.
-
Class IIofficial
Instrumentation Laboratory recalls Brand Name: GEM Premier 7000 with iQM3 Product Name: GEM Premier 7000 with iQM3 Model/Catalog Number: 00000015Potential for Bar Code Reader (BCR) wands, used with the GEM Premier 5000 and GEM Premier 7000 systems, to assign patient results to an incorrect patient ID.
-
Class IIofficial
Instrumentation Laboratory recalls Brand Name: GEM Premier 5000 Product Name: GEM Premier 5000 Model/Catalog Number: 00024019255 portable blood aPotential for Bar Code Reader (BCR) wands, used with the GEM Premier 5000 and GEM Premier 7000 systems, to assign patient results to an incorrect patient ID.
-
Class IIofficial
Instrumentation Laboratory recalls Brand Name: GEM Premier 5000 Product Name: GEM Premier 5000 Model/Catalog Number: 00024019255 portable blood aPotential for Bar Code Reader (BCR) wands, used with the GEM Premier 5000 and GEM Premier 7000 systems, to assign patient results to an incorrect patient ID.
-
Class IIofficial
Instrumentation Laboratory recalls Brand Name: GEM Premier 5000 Product Name: GEM Premier 5000 Model/Catalog Number: 00024019255 Portable blood aPotential for Bar Code Reader (BCR) wands, used with the GEM Premier 5000 and GEM Premier 7000 systems, to assign patient results to an incorrect patient ID.
-
Class IIofficial
Instrumentation Laboratory recalls Brand Name: GEM Premier 5000 Product Name: GEM Premier 5000 Model/Catalog Number: 00024019255 Portable blood aPotential for Bar Code Reader (BCR) wands, used with the GEM Premier 5000 and GEM Premier 7000 systems, to assign patient results to an incorrect patient ID.
-
Class Iofficial
Draeger, Inc. recalls Atlan A350. Model Number: 8211500. anesthesia workstationThe Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.
-
Class Iofficial
Draeger, Inc. recalls Atlan A350XL. Model Number: 8621600. anesthesia workstationThe Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.
-
Class IIofficial
Siemens Healthcare Diagnostics, Inc. recalls Siemens Dimension Tacrolimus Flex Reagent Cartridge (TAC)A potential positive bias in quality control (QC) and patient sample results in three lots of the Dimension Tacrolimus (TAC) Flex reagent cartridge.
-
Class IIofficial
Fresenius Medical Care Holdings, Inc. recalls Bicarby Dialysate; Model number: RFP-400-G;There is an error in the labeling of the GTIN on the product. When scanned; however, the case labeling, lot number, and printed labeling on the product correctly identifies the product code and the solution contained in the bag. In the event the incorrect GTIN labeling is used as the singular source for product identification, unrecognized use of a potassium solution with a concentration differing from the prescription may result.
-
Class IIofficial
Teleflex LLC recalls Brand Name: Pilling Wecksorb Product Name: Pilling Wecksorb Neurosponges (Cotton) Pilling¿ Wecksorb NeurospongDue to out-of-specification, or potential out-of-specification, levels of endotoxins
-
Class IIofficial
AVID Medical, Inc. recalls Halyard EYE PACK. Model/Catalog Number: LVEY27 Convenience kit.Incorrect concentration of Betadine solution contained within certain AVID convenience kits.
-
Class IIofficial
PIE Medical Imaging B.V. recalls 3mensio Workstation (Vascular Fenestrated) softwareUser measurement value reported by diagnostic bioimaging software intended to measure/visualize cardiovascular structures may be larger (0.8% to 34.4%) than the actual anatomical dimension, if all the following occur: affected software is used, and the user has access only to 3D Echo OR Mitral 3D Echo, and the dataset being analyzed is a Philips 3D echo dataset containing Color Doppler information
-
Class Iofficial
Fresenius Kabi USA, LLC recalls LVP software of the Ivenix Infusion System (IIS); Product Code LVP-SW-0005.Potential software anomaly that can cause the pump to report a false battery-health value (typically 69%) and trigger a battery-depletion alarm that will shut down the pump unless it is reconnected to AC mains power.
-
Class Iofficial
Fresenius Kabi USA, LLC recalls Ivenix Infusion System (IIS)Fresenius Kabi has been made aware via customer complaints that dropping or severe jarring of the Ivenix LVP may cause damage to internal components that may not be apparent or readily visible and the devices therefore remained in service. More specifically, drops or jarring on a hard surface can cause the touchscreen input cable to loosen, resulting in unintended touchscreen behavior on the LVP, including random screen touches and/or lack of screen responsiveness to user input. Under the reported failure condition, unintended touch inputs may register at arbitrary screen locations without user interaction. These inputs occur intermittently and without a predictable pattern.
-
Class Iofficial
B Braun Medical Inc recalls Spinal Needle procedure kits: Material Description (Material Number) 24GA. PENCAN SPINAL BUP. TRAY W/DURAPREPThese procedure kits contain Bupivacaine Hydrochloride in Dextrose Injection, USP. The manufacturer of the Bupivacaine, Huons Co, Ltd., has issued a voluntary recall due to potential quality issues identified during a recent FDA manufacturing facility inspection.
-
Class Iofficial
B Braun Medical Inc recalls Spinocan Spinal Needle procedure kits: Material Description (Material Number): Spinocan 26 Ga. x 3-1/2 in. (90These procedure kits contain Bupivacaine Hydrochloride in Dextrose Injection, USP. The manufacturer of the Bupivacaine, Huons Co, Ltd., has issued a voluntary recall due to potential quality issues identified during a recent FDA manufacturing facility inspection.
-
Class Iofficial
B Braun Medical Inc recalls Design Options Spinal Tray Spinal Needle Anesthesia procedure kits: Material Description (Material Number): SSThese procedure kits contain Bupivacaine Hydrochloride in Dextrose Injection, USP. The manufacturer of the Bupivacaine, Huons Co, Ltd., has issued a voluntary recall due to potential quality issues identified during a recent FDA manufacturing facility inspection.
-
Class IIofficial
Stryker Corporation recalls Neptune Smart Docking Station (120V)It has been identified that the impacted docking station (SN 2521008873) was not product released from manufacturing and may not have completed required testing prior to being shipped to your facility. There is a potential risk of electrical shock due to the device s safety testing being unverifiable.
-
Class Iofficial
Medline Industries, LP recalls Medline Convenience Kits: 1) DYNJ905503FThe Medline Convenience Kits contain Arrow Kits and Sets that include Lidocaine Hydrochloride Injection, USP and Bupivacaine Hydrochloride in Dextrose Injection, USP which were recalled due to quality issues identified during a recent FDA inspection of the supplier.
-
Class Iofficial
Medline Industries, LP recalls Medline Convenience Kits: 1) OPEN HEART CDSThe Medline Convenience Kits contain Arrow Kits and Sets that include Lidocaine Hydrochloride Injection, USP and Bupivacaine Hydrochloride in Dextrose Injection, USP which were recalled due to quality issues identified during a recent FDA inspection of the supplier.
-
Class Iofficial
Medline Industries, LP recalls Medline Convenience Kits: 1) DRAWER 6 CENTRAL LINE INSERTIOThe Medline Convenience Kits contain Arrow Kits and Sets that include Lidocaine Hydrochloride Injection, USP and Bupivacaine Hydrochloride in Dextrose Injection, USP which were recalled due to quality issues identified during a recent FDA inspection of the supplier.
-
Class Iofficial
Medline Industries, LP recalls Medline Convenience Kits: 1) DRAWER 6A ADULT CENTRAL LINEThe Medline Convenience Kits contain Arrow Kits and Sets that include Lidocaine Hydrochloride Injection, USP and Bupivacaine Hydrochloride in Dextrose Injection, USP which were recalled due to quality issues identified during a recent FDA inspection of the supplier.
-
Class IIofficial
Criticare Technologies, Inc. recalls nGenuity Patient Monitor with Capnography (ETCO2) CAT 8100E1-ST-USComplaint trend where "Checkline Error" alarm presents at device setup, rendering the CO2 monitoring function inoperable. May result in procedure delay.
-
Class IIofficial
IPG Medical Corporation recalls Thuvera Laser Console. Software Version 2.2. Portable Thulium Fiber Laser.Software anomaly that causes a false display of error code 5018.
-
Class IIofficial
Olympus Corporation of the Americas recalls Olympus Toomey Evacuator 30cc with Adaptor. Model Number: 188.Some sterilization parameters included in the IFU may not consistently render the devices adequately sterile and therefore should no longer be used to sterilize the devices.
-
Class IIofficial
Olympus Corporation of the Americas recalls Olympus Toomey Evacuator 50cc with Adaptor. Model Number: 188A.Some sterilization parameters included in the IFU may not consistently render the devices adequately sterile and therefore should no longer be used to sterilize the devices.
-
Class IIofficial
Olympus Corporation of the Americas recalls Olympus Ellik Evacuator Includes Glass BodySome sterilization parameters included in the IFU may not consistently render the devices adequately sterile and therefore should no longer be used to sterilize the devices.
-
Class IIofficial
Olympus Corporation of the Americas recalls Olympus Glass Body for 194 Ellik Evacuator. Model Number: 190A.Some sterilization parameters included in the IFU may not consistently render the devices adequately sterile and therefore should no longer be used to sterilize the devices.
-
Class Iofficial
Kayserbetten Gmbh & Co. Kg recalls Pediatric care bed; Product Designation: KayserBett IDA;If the adjustment functions of the hand control, as instructed in the Manual, are not locked with the provided key when the patient (child) is unattended, there is an increased risk that the patient (child) or third parties (children) will operate the hand control and trap themselves or others under the bed frame or between the bed frame and the floor. This can lead to serious injuries or even death of the patient (child) or third parties (children).
-
Class IIofficial
Micromed S.p.A. recalls SD LTM 64 PLUSNatus has become aware of two complaints related to the SD LTM 64 PLUS where the EEG traces acquired from two or four different amplifiers used in a multiple amplifier configuration (128 or 256 channels) running firmware version 2022.02 were displayed with a shift of 1 second between the 64channels blocks. In rare circumstances, a fixed 1-second temporal delay may appear between two 64-channel modules with firmware 2021.02, 2022.01 or 2022.02 in a multi-module SEEG configuration, potentially leading to misinterpretation of results.
-
Class IIofficial
Medline Industries, LP recalls MEDLINE Urethral CatheterDuring an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
-
Class IIofficial
Medline Industries, LP recalls MEDLINE Urethral Catheterization TrayDuring an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
-
Class IIofficial
Medline Industries, LP recalls MEDLINE Pre-connected Urethral Catheterization Tray and BagDuring an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
-
Class IIofficial
Medline Industries, LP recalls MEDLINE Urethral Catheterization Tray and BagDuring an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
-
Class IIofficial
Medline Industries, LP recalls MEDLINE Intermittent Latex Catheter Coude Pre-Connected to Collection BagDuring an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
-
Class IIofficial
Medline Industries, LP recalls MEDLINE Intermittent Latex Catheter Coude Pre-Connected to Collection BagDuring an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
-
Class IIofficial
Medline Industries, LP recalls MEDLINE Medical Procedure Kits labeled as: LAPAROSCOPY COMBINED - NO NDLDuring an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
-
Class IIofficial
Medline Industries, LP recalls MEDLINE Medical Procedure Kits labeled as: 1) BASIC HEART PACKDuring an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
-
Class IIofficial
Medline Industries, LP recalls MEDLINE Medical Procedure Kits labeled as CH OPEN HEARTDuring an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.